Efficient assay for total antioxidant capacity in human plasma using a 96-well microplte

Yasuhiro Kambayashi, Nguyen Thanh Binh, Hiroki W. Asakura, Yuri Hibino, Yoshiaki Hitomi, Hiroyuki Nakamura, Keiki Ogino

研究成果査読

89 被引用数 (Scopus)

抄録

In the present study, we tried to establish an efficient assay for total antioxidant capacity (TAC) in human plasma using a 96-well microplate. TAC was assessed using lag time by antioxidants against the myoglobin-induced oxidation of 2,2′-azino-di(3-ethylbenzthiazoline-6-sulfonic acid (ABTS) with hydrogen peroxide, and expressed as Trolox equivalent. The linearity of the calibration curve with Trolox was maintained with the Trolox concentration range from 2.5 μM to 25 μM (R2 = 0.997). The assay was applied to the measurement of TAC in healthy human plasma. Coefficient of variation in intraday assay was 2.4%. Difference was not observed in interday assay. Plasma TAC of men ((569 ± 41) μM Trolox equivalent; n = 6) was higher than that of women ((430 ± 28) μM Trolox equivalent; n = 4). After the vegetable juice was drunk for 1 week, the increase in plasma TAC was observed in almost all the volunteers. In summary, we developed the efficient assay for plasma TAC using a 96-well microplate.

本文言語English
ページ(範囲)46-51
ページ数6
ジャーナルjournal of clinical biochemistry and nutrition
44
1
DOI
出版ステータスPublished - 1月 2009
外部発表はい

ASJC Scopus subject areas

  • 医学(その他)
  • 栄養および糖尿病
  • 臨床生化学

フィンガープリント

「Efficient assay for total antioxidant capacity in human plasma using a 96-well microplte」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル