Development and evaluation of the simple answering device using under high magnetic field

Masayuki Kitazawa, Jing Long Wu, Kiyoshi Negoro, Masahiro Hashida, Junichi Ogasawara, Syuichi Yamauchi, Taizou Sanada, Naofumi Matsunaga, Yoshiro Sakai

研究成果査読

抄録

Magnetic resonance imaging (MRI) is a tomographical scanning system that has a strong magnetic field. In the MRI room, ferric materials and noisy electric circuits cannot be present or function At an experiment that investigates a human brain function using the fMRI (functional MRI), such as the phase discrimination of temporal frequency between visual and auditory, it is difficult to measure the answering times and its time quantitatively At this point of view, we have developed a new simple answering device. This device is made of the mouse that is popular at office. For the strong magnetic field, its circuit is removed except the clicking part. As it can be connected to a personal computer, we cart measure the answering times and its time during the experiment using the fMRI. To do so, there are some problems to solve, such as the strong magnetic field. In this paper, the performance of the answering device is demonstrated by several basic experiments to solve these problems.

本文言語English
ページ(範囲)3254-3259
ページ数6
ジャーナルNippon Kikai Gakkai Ronbunshu, C Hen/Transactions of the Japan Society of Mechanical Engineers, Part C
68
11
DOI
出版ステータスPublished - 11月 2002
外部発表はい

ASJC Scopus subject areas

  • 材料力学
  • 機械工学
  • 産業および生産工学

フィンガープリント

「Development and evaluation of the simple answering device using under high magnetic field」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル