A new navigation system for minimally invasive total knee arthroplasty.

Yusuke Yokoyama, Nobuhiro Abe, Kazuo Fujiwara, Masahiko Suzuki, Yoshikazu Nakajima, Naohiko Sugita, Mamoru Mitsuishi, Yoshio Nakashima, Toshifumi Ozaki

研究成果査読

5 被引用数 (Scopus)

抄録

A computer-assisted navigation system to be used for total knee arthroplasties (TKAs) was reported to improve the accuracy of bone resection and result in precise implant placement, but the concomitant surgical invasion and time consumption are clinical problems. We developed a computed tomography (CT)-based navigation system (NNS) to be used for minimally invasive TKA. It requires only the reference points from a small limited area of the medial femoral condyle and proximal tibia through a skin incision to obtain optical images. Here we evaluated the usefulness and accuracy of the NNS in comparison with the commercially available BrainLAB image-free navigation system (BLS). In a clinical experiment, the registration times obtained with the NNS tended to be shorter than those obtained with the BLS, but not significantly so. The NNS group tended to be in the extended position in the sagittal plane of the distal femur within 3 degrees, and the BLS group showed rather flexed deviation in the sagittal plane of the anterior femur.

本文言語English
ページ(範囲)351-358
ページ数8
ジャーナルActa medica Okayama
67
6
出版ステータスPublished - 2013

ASJC Scopus subject areas

  • 生化学、遺伝学、分子生物学(全般)

フィンガープリント

「A new navigation system for minimally invasive total knee arthroplasty.」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル