A caffeine-sensitive membrane electrode: Previous misleading report and present approach

Takashi Katsu, Yumi Tsunamoto, Nobumitsu Hanioka, Keiko Komagoe, Kazufumi Masuda, Shizuo Narimatsu

研究成果査読

19 被引用数 (Scopus)

抄録

Although a previous study [S.S.M. Hassan, M.A. Ahmed, M.M. Saoudi, Anal. Chem. 57 (1985) 1126] had shown that a caffeine-sensitive electrode made with picrylsulfonate and 1-octanol as a cation-exchanger and a solvent mediator, respectively, had a wide working pH range (5.5-9.5) and exhibited a Nernstian response, we could not find such response in this electrode. The present result was reasonable, because the pKa value of caffeinium ion was reported to be around 0.7 and the neutral form of caffeine was predominant in the pH range examined. Thus, we reinvestigated the response characteristics of a caffeine electrode, taking into consideration the pKa value, and constructed a new electrode with a combination of the lipophilic cation-exchanger, tetrakis[3,5-bis(2-methoxyhexafluoro-2-propyl)phenyl]borate (HFPB), and the solvent mediator with high degree of dielectric constant, 2-fluoro-2′-nitrodiphenyl ether (FNDPE). This electrode showed a pH-dependent response to caffeinium ion and gave a detection limit of 50 μM with a slope of 55 mV per concentration decade at pH 2. The use of other solvent mediators was less effective than that of FNDPE. The electrode was applied for the determination of caffeine in some central stimulants.

本文言語English
ページ(範囲)50-54
ページ数5
ジャーナルAnalytica Chimica Acta
620
1-2
DOI
出版ステータスPublished - 7月 14 2008
外部発表はい

ASJC Scopus subject areas

  • 分析化学
  • 生化学
  • 環境化学
  • 分光学

フィンガープリント

「A caffeine-sensitive membrane electrode: Previous misleading report and present approach」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル