Abstract
Significance: Polyp size is important for selecting the surveillance interval or treatment policy. Nevertheless, it is challenging to accurately estimate the polyp size during endoscopy. An easy and cost-effective function to assist in polyp size estimation is required. Aim: To propose a virtual scale function for endoscopy and evaluate its performance and expected accuracy. Approach: An adaptive virtual scale behavior was demonstrated. The measurement error of the virtual scale along the distance between the tip of the endoscope and the object plane was evaluated using graph paper. The accuracy of polyp size estimation by an expert endoscopist was compared with the accuracy of the biopsy forceps method using phantom images. Results: The measurement errors of the virtual scale were ≤ 0.7 mm when the distance to the graph paper, which faced the tip of the endoscope, varied from 4 to 30 mm. The accuracy with the virtual scale was significantly higher than that obtained with biopsy forceps (5.3 ± 5.5 % versus 11.9 ± 9.4 %, P < 0.001). Conclusions: The virtual scale function, which operates in real-time without any additional device, can be used to estimate polyp sizes easily and accurately with endoscopy.
Original language | English |
---|---|
Article number | 096002 |
Journal | Journal of Biomedical Optics |
Volume | 26 |
Issue number | 9 |
DOIs | |
Publication status | Published - Sept 1 2021 |
Keywords
- endoscopy
- image processing
- laser applications
- real-time imaging
- size estimation
- virtual scale
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Atomic and Molecular Physics, and Optics
- Biomaterials
- Biomedical Engineering