SU‐E‐T‐106: Evaluation of Sensitivity and Uniformity of New Radiochromic Film with Two Commercial Scanners

T. Kamomae, Y. Miyabe, A. Sawada, Masataka Oita, T. Fujimoto, S. Yano, M. Nakata, S. Nakayama, M. Hiraoka

Research output: Contribution to journalArticle

Abstract

Purpose: A newly introduced radiochromic film, the GAFCHROMIC EBT3, has been expected as much useful device for the IMRT dosimetry. The purpose of this study was to investigate the sensitivity and the uniformity of the films between an Epson ES‐10000G flatbed scanner and a Vidar DosimetryPRO Advantage (Red) scanner. Methods: Doses ranging from 1 cGy to 1600 cGy with 15‐MV photon beam was irradiated to the film in a solid water phantom, respectively. All of the films were then digitized after irradiation using both two scanners. Sensitivities, local fluctuations of the film with two scanners were evaluated. Local fluctuations were defined as the relative (percent) standard deviation of the film response in ROIs (3 cmx3 cm). Results: As to the Vidar scanner, the sensitivity of the film was higher for low dose range (below <400 cGy). While, as to the Epson scanner, the sensitivity using the red color channel was higher than others for low dose range. At high dose range (above >400 cGy), the green color channel had higher sensitivity than others. The Vidar scanner exhibited the lower local fluctuations than the Epson scanner for all dose ranges. For the Epson scanner, the red color channel had the lower local fluctuations than the green and blue color channel for all dose ranges. Conclusions: This study shows the characteristics of the new EBT3 films, in conjunction with the Epson ES‐10000G flatbed scanner and the Vidar DosimetryPRO Advantage (Red) scanner.

Original languageEnglish
Pages (from-to)3727
Number of pages1
JournalMedical Physics
Volume39
Issue number6
DOIs
Publication statusPublished - 2012

Fingerprint

Color
Photons
Equipment and Supplies
Water

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐E‐T‐106 : Evaluation of Sensitivity and Uniformity of New Radiochromic Film with Two Commercial Scanners. / Kamomae, T.; Miyabe, Y.; Sawada, A.; Oita, Masataka; Fujimoto, T.; Yano, S.; Nakata, M.; Nakayama, S.; Hiraoka, M.

In: Medical Physics, Vol. 39, No. 6, 2012, p. 3727.

Research output: Contribution to journalArticle

Kamomae, T, Miyabe, Y, Sawada, A, Oita, M, Fujimoto, T, Yano, S, Nakata, M, Nakayama, S & Hiraoka, M 2012, 'SU‐E‐T‐106: Evaluation of Sensitivity and Uniformity of New Radiochromic Film with Two Commercial Scanners', Medical Physics, vol. 39, no. 6, pp. 3727. https://doi.org/10.1118/1.4735164
Kamomae, T. ; Miyabe, Y. ; Sawada, A. ; Oita, Masataka ; Fujimoto, T. ; Yano, S. ; Nakata, M. ; Nakayama, S. ; Hiraoka, M. / SU‐E‐T‐106 : Evaluation of Sensitivity and Uniformity of New Radiochromic Film with Two Commercial Scanners. In: Medical Physics. 2012 ; Vol. 39, No. 6. pp. 3727.
@article{11a964b701494d409556d420e4ae59f7,
title = "SU‐E‐T‐106: Evaluation of Sensitivity and Uniformity of New Radiochromic Film with Two Commercial Scanners",
abstract = "Purpose: A newly introduced radiochromic film, the GAFCHROMIC EBT3, has been expected as much useful device for the IMRT dosimetry. The purpose of this study was to investigate the sensitivity and the uniformity of the films between an Epson ES‐10000G flatbed scanner and a Vidar DosimetryPRO Advantage (Red) scanner. Methods: Doses ranging from 1 cGy to 1600 cGy with 15‐MV photon beam was irradiated to the film in a solid water phantom, respectively. All of the films were then digitized after irradiation using both two scanners. Sensitivities, local fluctuations of the film with two scanners were evaluated. Local fluctuations were defined as the relative (percent) standard deviation of the film response in ROIs (3 cmx3 cm). Results: As to the Vidar scanner, the sensitivity of the film was higher for low dose range (below <400 cGy). While, as to the Epson scanner, the sensitivity using the red color channel was higher than others for low dose range. At high dose range (above >400 cGy), the green color channel had higher sensitivity than others. The Vidar scanner exhibited the lower local fluctuations than the Epson scanner for all dose ranges. For the Epson scanner, the red color channel had the lower local fluctuations than the green and blue color channel for all dose ranges. Conclusions: This study shows the characteristics of the new EBT3 films, in conjunction with the Epson ES‐10000G flatbed scanner and the Vidar DosimetryPRO Advantage (Red) scanner.",
author = "T. Kamomae and Y. Miyabe and A. Sawada and Masataka Oita and T. Fujimoto and S. Yano and M. Nakata and S. Nakayama and M. Hiraoka",
year = "2012",
doi = "10.1118/1.4735164",
language = "English",
volume = "39",
pages = "3727",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐E‐T‐106

T2 - Evaluation of Sensitivity and Uniformity of New Radiochromic Film with Two Commercial Scanners

AU - Kamomae, T.

AU - Miyabe, Y.

AU - Sawada, A.

AU - Oita, Masataka

AU - Fujimoto, T.

AU - Yano, S.

AU - Nakata, M.

AU - Nakayama, S.

AU - Hiraoka, M.

PY - 2012

Y1 - 2012

N2 - Purpose: A newly introduced radiochromic film, the GAFCHROMIC EBT3, has been expected as much useful device for the IMRT dosimetry. The purpose of this study was to investigate the sensitivity and the uniformity of the films between an Epson ES‐10000G flatbed scanner and a Vidar DosimetryPRO Advantage (Red) scanner. Methods: Doses ranging from 1 cGy to 1600 cGy with 15‐MV photon beam was irradiated to the film in a solid water phantom, respectively. All of the films were then digitized after irradiation using both two scanners. Sensitivities, local fluctuations of the film with two scanners were evaluated. Local fluctuations were defined as the relative (percent) standard deviation of the film response in ROIs (3 cmx3 cm). Results: As to the Vidar scanner, the sensitivity of the film was higher for low dose range (below <400 cGy). While, as to the Epson scanner, the sensitivity using the red color channel was higher than others for low dose range. At high dose range (above >400 cGy), the green color channel had higher sensitivity than others. The Vidar scanner exhibited the lower local fluctuations than the Epson scanner for all dose ranges. For the Epson scanner, the red color channel had the lower local fluctuations than the green and blue color channel for all dose ranges. Conclusions: This study shows the characteristics of the new EBT3 films, in conjunction with the Epson ES‐10000G flatbed scanner and the Vidar DosimetryPRO Advantage (Red) scanner.

AB - Purpose: A newly introduced radiochromic film, the GAFCHROMIC EBT3, has been expected as much useful device for the IMRT dosimetry. The purpose of this study was to investigate the sensitivity and the uniformity of the films between an Epson ES‐10000G flatbed scanner and a Vidar DosimetryPRO Advantage (Red) scanner. Methods: Doses ranging from 1 cGy to 1600 cGy with 15‐MV photon beam was irradiated to the film in a solid water phantom, respectively. All of the films were then digitized after irradiation using both two scanners. Sensitivities, local fluctuations of the film with two scanners were evaluated. Local fluctuations were defined as the relative (percent) standard deviation of the film response in ROIs (3 cmx3 cm). Results: As to the Vidar scanner, the sensitivity of the film was higher for low dose range (below <400 cGy). While, as to the Epson scanner, the sensitivity using the red color channel was higher than others for low dose range. At high dose range (above >400 cGy), the green color channel had higher sensitivity than others. The Vidar scanner exhibited the lower local fluctuations than the Epson scanner for all dose ranges. For the Epson scanner, the red color channel had the lower local fluctuations than the green and blue color channel for all dose ranges. Conclusions: This study shows the characteristics of the new EBT3 films, in conjunction with the Epson ES‐10000G flatbed scanner and the Vidar DosimetryPRO Advantage (Red) scanner.

UR - http://www.scopus.com/inward/record.url?scp=85024820328&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024820328&partnerID=8YFLogxK

U2 - 10.1118/1.4735164

DO - 10.1118/1.4735164

M3 - Article

AN - SCOPUS:85024820328

VL - 39

SP - 3727

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -