Shift-and-add tomosynthesis of a finger joint by X-ray dark-field imaging: Difference due to tomographic angle

Daisuke Shimao, Toshiyuki Kunisada, Hiroshi Sugiyama, Masami Ando

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

A tomogram of a finger joint showing articular cartilage was generated based on X-ray dark-field imaging (XDFI) using the shift-and-add tomosynthesis algorithm. The experiment was performed at beamline 14B of the Photon Factory in Tsukuba, Japan, using synchrotron X-rays from a vertical wiggler. The incident X-ray energy was 36.0 keV. The X-ray optics for XDFI comprised two Si crystals: an asymmetric cut Si (2 2 0) monochromator-collimator and a 1.1-mm thick Si (2 2 0) Laue-case analyzer. The object was an intact cadaveric proximal interphalangeal joint fixed in formalin. Raw projection data were acquired by XDFI in a total of 41 views through an angle of 20° in 0.5° increments. The object and detector were synchronously rotated such that the fulcrum plane in the object and detector plane remained parallel. The X-ray dose for one piece of raw projection data was set to one-eleventh of that for one standard projection image by XDFI. Eleven views through an angle of 10° in increments of 1° of all 41 appropriately shifted raw projection data were added to produce arbitrary tomograms parallel to the fulcrum plane. We obtained a clear tomogram of the finger joint including the articular cartilage with the moderate artifact peculiar to tomosynthesis. Consequently, arbitrary tomograms can be obtained for the same X-ray dose as that received for one standard projection image by XDFI. The fact that an inner structure such as articular cartilage, which is invisible to conventional X-ray imaging methods, has been visualized on a tomogram with preserved refraction-enhanced contrast, is of considerable significance to clinical medicine.

Original languageEnglish
Pages (from-to)27-31
Number of pages5
JournalEuropean Journal of Radiology
Volume68
Issue number3 SUPPL.
DOIs
Publication statusPublished - Dec 2008

Fingerprint

Finger Joint
X-Rays
Articular Cartilage
Synchrotrons
Clinical Medicine
Photons
Artifacts
Formaldehyde
Japan
Joints

Keywords

  • Cartilage
  • Laue analyzer
  • Shift-and-add
  • Synchrotron radiation
  • Tomosynthesis
  • X-ray dark field imaging

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Shift-and-add tomosynthesis of a finger joint by X-ray dark-field imaging : Difference due to tomographic angle. / Shimao, Daisuke; Kunisada, Toshiyuki; Sugiyama, Hiroshi; Ando, Masami.

In: European Journal of Radiology, Vol. 68, No. 3 SUPPL., 12.2008, p. 27-31.

Research output: Contribution to journalArticle

@article{03fc2508dd5442aba871f7105e4f216f,
title = "Shift-and-add tomosynthesis of a finger joint by X-ray dark-field imaging: Difference due to tomographic angle",
abstract = "A tomogram of a finger joint showing articular cartilage was generated based on X-ray dark-field imaging (XDFI) using the shift-and-add tomosynthesis algorithm. The experiment was performed at beamline 14B of the Photon Factory in Tsukuba, Japan, using synchrotron X-rays from a vertical wiggler. The incident X-ray energy was 36.0 keV. The X-ray optics for XDFI comprised two Si crystals: an asymmetric cut Si (2 2 0) monochromator-collimator and a 1.1-mm thick Si (2 2 0) Laue-case analyzer. The object was an intact cadaveric proximal interphalangeal joint fixed in formalin. Raw projection data were acquired by XDFI in a total of 41 views through an angle of 20° in 0.5° increments. The object and detector were synchronously rotated such that the fulcrum plane in the object and detector plane remained parallel. The X-ray dose for one piece of raw projection data was set to one-eleventh of that for one standard projection image by XDFI. Eleven views through an angle of 10° in increments of 1° of all 41 appropriately shifted raw projection data were added to produce arbitrary tomograms parallel to the fulcrum plane. We obtained a clear tomogram of the finger joint including the articular cartilage with the moderate artifact peculiar to tomosynthesis. Consequently, arbitrary tomograms can be obtained for the same X-ray dose as that received for one standard projection image by XDFI. The fact that an inner structure such as articular cartilage, which is invisible to conventional X-ray imaging methods, has been visualized on a tomogram with preserved refraction-enhanced contrast, is of considerable significance to clinical medicine.",
keywords = "Cartilage, Laue analyzer, Shift-and-add, Synchrotron radiation, Tomosynthesis, X-ray dark field imaging",
author = "Daisuke Shimao and Toshiyuki Kunisada and Hiroshi Sugiyama and Masami Ando",
year = "2008",
month = "12",
doi = "10.1016/j.ejrad.2008.04.037",
language = "English",
volume = "68",
pages = "27--31",
journal = "Journal of Medical Imaging",
issn = "0720-048X",
publisher = "Elsevier Ireland Ltd",
number = "3 SUPPL.",

}

TY - JOUR

T1 - Shift-and-add tomosynthesis of a finger joint by X-ray dark-field imaging

T2 - Difference due to tomographic angle

AU - Shimao, Daisuke

AU - Kunisada, Toshiyuki

AU - Sugiyama, Hiroshi

AU - Ando, Masami

PY - 2008/12

Y1 - 2008/12

N2 - A tomogram of a finger joint showing articular cartilage was generated based on X-ray dark-field imaging (XDFI) using the shift-and-add tomosynthesis algorithm. The experiment was performed at beamline 14B of the Photon Factory in Tsukuba, Japan, using synchrotron X-rays from a vertical wiggler. The incident X-ray energy was 36.0 keV. The X-ray optics for XDFI comprised two Si crystals: an asymmetric cut Si (2 2 0) monochromator-collimator and a 1.1-mm thick Si (2 2 0) Laue-case analyzer. The object was an intact cadaveric proximal interphalangeal joint fixed in formalin. Raw projection data were acquired by XDFI in a total of 41 views through an angle of 20° in 0.5° increments. The object and detector were synchronously rotated such that the fulcrum plane in the object and detector plane remained parallel. The X-ray dose for one piece of raw projection data was set to one-eleventh of that for one standard projection image by XDFI. Eleven views through an angle of 10° in increments of 1° of all 41 appropriately shifted raw projection data were added to produce arbitrary tomograms parallel to the fulcrum plane. We obtained a clear tomogram of the finger joint including the articular cartilage with the moderate artifact peculiar to tomosynthesis. Consequently, arbitrary tomograms can be obtained for the same X-ray dose as that received for one standard projection image by XDFI. The fact that an inner structure such as articular cartilage, which is invisible to conventional X-ray imaging methods, has been visualized on a tomogram with preserved refraction-enhanced contrast, is of considerable significance to clinical medicine.

AB - A tomogram of a finger joint showing articular cartilage was generated based on X-ray dark-field imaging (XDFI) using the shift-and-add tomosynthesis algorithm. The experiment was performed at beamline 14B of the Photon Factory in Tsukuba, Japan, using synchrotron X-rays from a vertical wiggler. The incident X-ray energy was 36.0 keV. The X-ray optics for XDFI comprised two Si crystals: an asymmetric cut Si (2 2 0) monochromator-collimator and a 1.1-mm thick Si (2 2 0) Laue-case analyzer. The object was an intact cadaveric proximal interphalangeal joint fixed in formalin. Raw projection data were acquired by XDFI in a total of 41 views through an angle of 20° in 0.5° increments. The object and detector were synchronously rotated such that the fulcrum plane in the object and detector plane remained parallel. The X-ray dose for one piece of raw projection data was set to one-eleventh of that for one standard projection image by XDFI. Eleven views through an angle of 10° in increments of 1° of all 41 appropriately shifted raw projection data were added to produce arbitrary tomograms parallel to the fulcrum plane. We obtained a clear tomogram of the finger joint including the articular cartilage with the moderate artifact peculiar to tomosynthesis. Consequently, arbitrary tomograms can be obtained for the same X-ray dose as that received for one standard projection image by XDFI. The fact that an inner structure such as articular cartilage, which is invisible to conventional X-ray imaging methods, has been visualized on a tomogram with preserved refraction-enhanced contrast, is of considerable significance to clinical medicine.

KW - Cartilage

KW - Laue analyzer

KW - Shift-and-add

KW - Synchrotron radiation

KW - Tomosynthesis

KW - X-ray dark field imaging

UR - http://www.scopus.com/inward/record.url?scp=56549091661&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=56549091661&partnerID=8YFLogxK

U2 - 10.1016/j.ejrad.2008.04.037

DO - 10.1016/j.ejrad.2008.04.037

M3 - Article

C2 - 18599237

AN - SCOPUS:56549091661

VL - 68

SP - 27

EP - 31

JO - Journal of Medical Imaging

JF - Journal of Medical Imaging

SN - 0720-048X

IS - 3 SUPPL.

ER -