Sheath-flow electrochemical detection of amino acids with a copper wire electrode in capillary electrophoresis

Junji Inoue, Takashi Kaneta, Totaro Imasaka

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Here, we report the detection of native amino acids using a sheath-flow electrochemical detector with a working electrode made of copper wire. A separation capillary that was inserted into a platinum tube in the detector acted as a grounded electrode for electrophoresis and as a flow channel for sheath liquid. Sheath liquid flowed outside the capillary to support the transport of the separated analytes to the working electrode for electrochemical detection. The copper wire electrode was aligned at the outlet of the capillary in a wall-jet configuration. Amino acids injected into the capillary were separated following elution from the end of the capillary and detection by the copper electrode. Three kinds of copper electrodes with different diameters-50, 125, and 300 μm-were examined to investigate the effect of the electrode diameter on sensitivity. The peak widths of the analytes were independent of the diameter of the working electrode, while the 300-μm electrode led to a decrease in the signal-to-noise ratio compared with the 50- and 125-μm electrodes, which showed no significant difference. The flow rate of the sheath liquid was also varied to optimize the detection conditions. The limits of detection for amino acids ranged from 4.4 to 27 μM under optimal conditions.

Original languageEnglish
Pages (from-to)2743-2747
Number of pages5
JournalElectrophoresis
Volume33
Issue number17
DOIs
Publication statusPublished - Sep 2012

Fingerprint

Capillary electrophoresis
Capillary Electrophoresis
Copper
Electrodes
Wire
Amino Acids
Liquids
Detectors
Signal-To-Noise Ratio
Channel flow
Platinum
Electrophoresis
Limit of Detection
Signal to noise ratio
Flow rate

Keywords

  • Amino acids
  • Amperometry
  • Capillary electrophoresis
  • Electrochemical detection
  • Sheath flow

ASJC Scopus subject areas

  • Biochemistry
  • Clinical Biochemistry

Cite this

Sheath-flow electrochemical detection of amino acids with a copper wire electrode in capillary electrophoresis. / Inoue, Junji; Kaneta, Takashi; Imasaka, Totaro.

In: Electrophoresis, Vol. 33, No. 17, 09.2012, p. 2743-2747.

Research output: Contribution to journalArticle

@article{8037b59dae8b4aad8bfd878ee070d0a4,
title = "Sheath-flow electrochemical detection of amino acids with a copper wire electrode in capillary electrophoresis",
abstract = "Here, we report the detection of native amino acids using a sheath-flow electrochemical detector with a working electrode made of copper wire. A separation capillary that was inserted into a platinum tube in the detector acted as a grounded electrode for electrophoresis and as a flow channel for sheath liquid. Sheath liquid flowed outside the capillary to support the transport of the separated analytes to the working electrode for electrochemical detection. The copper wire electrode was aligned at the outlet of the capillary in a wall-jet configuration. Amino acids injected into the capillary were separated following elution from the end of the capillary and detection by the copper electrode. Three kinds of copper electrodes with different diameters-50, 125, and 300 μm-were examined to investigate the effect of the electrode diameter on sensitivity. The peak widths of the analytes were independent of the diameter of the working electrode, while the 300-μm electrode led to a decrease in the signal-to-noise ratio compared with the 50- and 125-μm electrodes, which showed no significant difference. The flow rate of the sheath liquid was also varied to optimize the detection conditions. The limits of detection for amino acids ranged from 4.4 to 27 μM under optimal conditions.",
keywords = "Amino acids, Amperometry, Capillary electrophoresis, Electrochemical detection, Sheath flow",
author = "Junji Inoue and Takashi Kaneta and Totaro Imasaka",
year = "2012",
month = "9",
doi = "10.1002/elps.201200059",
language = "English",
volume = "33",
pages = "2743--2747",
journal = "Electrophoresis",
issn = "0173-0835",
publisher = "Wiley-VCH Verlag",
number = "17",

}

TY - JOUR

T1 - Sheath-flow electrochemical detection of amino acids with a copper wire electrode in capillary electrophoresis

AU - Inoue, Junji

AU - Kaneta, Takashi

AU - Imasaka, Totaro

PY - 2012/9

Y1 - 2012/9

N2 - Here, we report the detection of native amino acids using a sheath-flow electrochemical detector with a working electrode made of copper wire. A separation capillary that was inserted into a platinum tube in the detector acted as a grounded electrode for electrophoresis and as a flow channel for sheath liquid. Sheath liquid flowed outside the capillary to support the transport of the separated analytes to the working electrode for electrochemical detection. The copper wire electrode was aligned at the outlet of the capillary in a wall-jet configuration. Amino acids injected into the capillary were separated following elution from the end of the capillary and detection by the copper electrode. Three kinds of copper electrodes with different diameters-50, 125, and 300 μm-were examined to investigate the effect of the electrode diameter on sensitivity. The peak widths of the analytes were independent of the diameter of the working electrode, while the 300-μm electrode led to a decrease in the signal-to-noise ratio compared with the 50- and 125-μm electrodes, which showed no significant difference. The flow rate of the sheath liquid was also varied to optimize the detection conditions. The limits of detection for amino acids ranged from 4.4 to 27 μM under optimal conditions.

AB - Here, we report the detection of native amino acids using a sheath-flow electrochemical detector with a working electrode made of copper wire. A separation capillary that was inserted into a platinum tube in the detector acted as a grounded electrode for electrophoresis and as a flow channel for sheath liquid. Sheath liquid flowed outside the capillary to support the transport of the separated analytes to the working electrode for electrochemical detection. The copper wire electrode was aligned at the outlet of the capillary in a wall-jet configuration. Amino acids injected into the capillary were separated following elution from the end of the capillary and detection by the copper electrode. Three kinds of copper electrodes with different diameters-50, 125, and 300 μm-were examined to investigate the effect of the electrode diameter on sensitivity. The peak widths of the analytes were independent of the diameter of the working electrode, while the 300-μm electrode led to a decrease in the signal-to-noise ratio compared with the 50- and 125-μm electrodes, which showed no significant difference. The flow rate of the sheath liquid was also varied to optimize the detection conditions. The limits of detection for amino acids ranged from 4.4 to 27 μM under optimal conditions.

KW - Amino acids

KW - Amperometry

KW - Capillary electrophoresis

KW - Electrochemical detection

KW - Sheath flow

UR - http://www.scopus.com/inward/record.url?scp=84866040191&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84866040191&partnerID=8YFLogxK

U2 - 10.1002/elps.201200059

DO - 10.1002/elps.201200059

M3 - Article

C2 - 22965720

AN - SCOPUS:84866040191

VL - 33

SP - 2743

EP - 2747

JO - Electrophoresis