In general, patients with recent versions of cochlear implants(CIs)can safely undergo Magnetic Resonance Imaging (MRI) within 1.5 T scanners, without the need of inner magnet removal surgery. However, even if they wear tight head bandages on their inner coil, complications can occur in CI users during MRI, such as pain, skin reaction, demagnetization of the internal magnet, and dislocation of the magnet out of its silicon flange. We encountered two cases of reversal of the inner magnet of a CI during MRI examination. In both cases, the inner magnets were reversed by 180 degrees, although they remained in the silicon flange. We ordered a polar-changed extra magnet for replacing the reversed inner magnet in one case, and performed a re-implant surgery of the inner magnet in the other. In recent years, the number of hearing-impaired patients with CIs has been increasing along with expansion of the medical indications for CI in the rapidly aging population of the developed countries. On the other hand, MRI is an indispensable clinical diagnostic tool in modern medical practice. Physicians and paramedics must be aware of possible complications that can occur during MRI in patients with CIs. Prompt assessment and appropriate treatment are important in patients with CIs developing complications during MRI examination.

Original languageEnglish
Pages (from-to)727-732
Number of pages6
JournalJournal of Otolaryngology of Japan
Volume120
Issue number5
Publication statusPublished - 2017

Fingerprint

Cochlear Implants
Magnets
Magnetic Resonance Imaging
Silicon
Allied Health Personnel
Bandages
Developed Countries
Hearing
Head
Physicians
Pain
Skin

Keywords

  • Cochlear implant
  • Complication of MRI
  • Magnet reversal

ASJC Scopus subject areas

  • Otorhinolaryngology
  • Medicine(all)

Cite this

Reversal of the inner magnet of a cochlear implant after magnet resonance imaging. / Kataoka, Yuko; Naito, Tomoyuki; Kariya, Shin; Sugaya, Akiko; Maeda, Yukihide; Fukushima, Kunihiro; Nishizaki, Kazunori.

In: Journal of Otolaryngology of Japan, Vol. 120, No. 5, 2017, p. 727-732.

Research output: Contribution to journalArticle

@article{32b72de15189476b8e4929f495020e42,
title = "Reversal of the inner magnet of a cochlear implant after magnet resonance imaging",
abstract = "In general, patients with recent versions of cochlear implants(CIs)can safely undergo Magnetic Resonance Imaging (MRI) within 1.5 T scanners, without the need of inner magnet removal surgery. However, even if they wear tight head bandages on their inner coil, complications can occur in CI users during MRI, such as pain, skin reaction, demagnetization of the internal magnet, and dislocation of the magnet out of its silicon flange. We encountered two cases of reversal of the inner magnet of a CI during MRI examination. In both cases, the inner magnets were reversed by 180 degrees, although they remained in the silicon flange. We ordered a polar-changed extra magnet for replacing the reversed inner magnet in one case, and performed a re-implant surgery of the inner magnet in the other. In recent years, the number of hearing-impaired patients with CIs has been increasing along with expansion of the medical indications for CI in the rapidly aging population of the developed countries. On the other hand, MRI is an indispensable clinical diagnostic tool in modern medical practice. Physicians and paramedics must be aware of possible complications that can occur during MRI in patients with CIs. Prompt assessment and appropriate treatment are important in patients with CIs developing complications during MRI examination.",
keywords = "Cochlear implant, Complication of MRI, Magnet reversal",
author = "Yuko Kataoka and Tomoyuki Naito and Shin Kariya and Akiko Sugaya and Yukihide Maeda and Kunihiro Fukushima and Kazunori Nishizaki",
year = "2017",
language = "English",
volume = "120",
pages = "727--732",
journal = "Journal of Otolaryngology of Japan",
issn = "0030-6622",
publisher = "Oto-Rhino-Laryngological Society of Japan Inc.",
number = "5",

}

TY - JOUR

T1 - Reversal of the inner magnet of a cochlear implant after magnet resonance imaging

AU - Kataoka, Yuko

AU - Naito, Tomoyuki

AU - Kariya, Shin

AU - Sugaya, Akiko

AU - Maeda, Yukihide

AU - Fukushima, Kunihiro

AU - Nishizaki, Kazunori

PY - 2017

Y1 - 2017

N2 - In general, patients with recent versions of cochlear implants(CIs)can safely undergo Magnetic Resonance Imaging (MRI) within 1.5 T scanners, without the need of inner magnet removal surgery. However, even if they wear tight head bandages on their inner coil, complications can occur in CI users during MRI, such as pain, skin reaction, demagnetization of the internal magnet, and dislocation of the magnet out of its silicon flange. We encountered two cases of reversal of the inner magnet of a CI during MRI examination. In both cases, the inner magnets were reversed by 180 degrees, although they remained in the silicon flange. We ordered a polar-changed extra magnet for replacing the reversed inner magnet in one case, and performed a re-implant surgery of the inner magnet in the other. In recent years, the number of hearing-impaired patients with CIs has been increasing along with expansion of the medical indications for CI in the rapidly aging population of the developed countries. On the other hand, MRI is an indispensable clinical diagnostic tool in modern medical practice. Physicians and paramedics must be aware of possible complications that can occur during MRI in patients with CIs. Prompt assessment and appropriate treatment are important in patients with CIs developing complications during MRI examination.

AB - In general, patients with recent versions of cochlear implants(CIs)can safely undergo Magnetic Resonance Imaging (MRI) within 1.5 T scanners, without the need of inner magnet removal surgery. However, even if they wear tight head bandages on their inner coil, complications can occur in CI users during MRI, such as pain, skin reaction, demagnetization of the internal magnet, and dislocation of the magnet out of its silicon flange. We encountered two cases of reversal of the inner magnet of a CI during MRI examination. In both cases, the inner magnets were reversed by 180 degrees, although they remained in the silicon flange. We ordered a polar-changed extra magnet for replacing the reversed inner magnet in one case, and performed a re-implant surgery of the inner magnet in the other. In recent years, the number of hearing-impaired patients with CIs has been increasing along with expansion of the medical indications for CI in the rapidly aging population of the developed countries. On the other hand, MRI is an indispensable clinical diagnostic tool in modern medical practice. Physicians and paramedics must be aware of possible complications that can occur during MRI in patients with CIs. Prompt assessment and appropriate treatment are important in patients with CIs developing complications during MRI examination.

KW - Cochlear implant

KW - Complication of MRI

KW - Magnet reversal

UR - http://www.scopus.com/inward/record.url?scp=85027051839&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85027051839&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:85027051839

VL - 120

SP - 727

EP - 732

JO - Journal of Otolaryngology of Japan

JF - Journal of Otolaryngology of Japan

SN - 0030-6622

IS - 5

ER -