Mean lung pressure during adult high-frequency oscillatory ventilation: An experimental study using a lung model

Takahiro Hirayama, Osamu Nagano, Naoki Shiba, Tetsuya Yumoto, Keiji Sato, Michihisa Terado, Toyomu Ugawa, Shingo Ichiba, Yoshihito Ujike

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.

Original languageEnglish
Pages (from-to)323-329
Number of pages7
JournalActa Medica Okayama
Volume68
Issue number6
Publication statusPublished - 2014

Fingerprint

High-Frequency Ventilation
Ventilation
Pressure
Lung
Mechanical Ventilators
Plastic containers
Resistors
Stroke Volume
Plastics
Compliance

Keywords

  • HFOV
  • Mean airway pressure
  • Mean lung pressure

ASJC Scopus subject areas

  • Medicine(all)
  • Biochemistry, Genetics and Molecular Biology(all)

Cite this

Hirayama, T., Nagano, O., Shiba, N., Yumoto, T., Sato, K., Terado, M., ... Ujike, Y. (2014). Mean lung pressure during adult high-frequency oscillatory ventilation: An experimental study using a lung model. Acta Medica Okayama, 68(6), 323-329.

Mean lung pressure during adult high-frequency oscillatory ventilation : An experimental study using a lung model. / Hirayama, Takahiro; Nagano, Osamu; Shiba, Naoki; Yumoto, Tetsuya; Sato, Keiji; Terado, Michihisa; Ugawa, Toyomu; Ichiba, Shingo; Ujike, Yoshihito.

In: Acta Medica Okayama, Vol. 68, No. 6, 2014, p. 323-329.

Research output: Contribution to journalArticle

Hirayama, T, Nagano, O, Shiba, N, Yumoto, T, Sato, K, Terado, M, Ugawa, T, Ichiba, S & Ujike, Y 2014, 'Mean lung pressure during adult high-frequency oscillatory ventilation: An experimental study using a lung model', Acta Medica Okayama, vol. 68, no. 6, pp. 323-329.
Hirayama T, Nagano O, Shiba N, Yumoto T, Sato K, Terado M et al. Mean lung pressure during adult high-frequency oscillatory ventilation: An experimental study using a lung model. Acta Medica Okayama. 2014;68(6):323-329.
Hirayama, Takahiro ; Nagano, Osamu ; Shiba, Naoki ; Yumoto, Tetsuya ; Sato, Keiji ; Terado, Michihisa ; Ugawa, Toyomu ; Ichiba, Shingo ; Ujike, Yoshihito. / Mean lung pressure during adult high-frequency oscillatory ventilation : An experimental study using a lung model. In: Acta Medica Okayama. 2014 ; Vol. 68, No. 6. pp. 323-329.
@article{8141bfed721444c09d3ffff1cea5ee79,
title = "Mean lung pressure during adult high-frequency oscillatory ventilation: An experimental study using a lung model",
abstract = "In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and {\%} of inspiratory time (IT), 50{\%} for R100, 33{\%} or 50{\%} for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.",
keywords = "HFOV, Mean airway pressure, Mean lung pressure",
author = "Takahiro Hirayama and Osamu Nagano and Naoki Shiba and Tetsuya Yumoto and Keiji Sato and Michihisa Terado and Toyomu Ugawa and Shingo Ichiba and Yoshihito Ujike",
year = "2014",
language = "English",
volume = "68",
pages = "323--329",
journal = "Acta Medica Okayama",
issn = "0386-300X",
publisher = "Okayama University",
number = "6",

}

TY - JOUR

T1 - Mean lung pressure during adult high-frequency oscillatory ventilation

T2 - An experimental study using a lung model

AU - Hirayama, Takahiro

AU - Nagano, Osamu

AU - Shiba, Naoki

AU - Yumoto, Tetsuya

AU - Sato, Keiji

AU - Terado, Michihisa

AU - Ugawa, Toyomu

AU - Ichiba, Shingo

AU - Ujike, Yoshihito

PY - 2014

Y1 - 2014

N2 - In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.

AB - In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.

KW - HFOV

KW - Mean airway pressure

KW - Mean lung pressure

UR - http://www.scopus.com/inward/record.url?scp=84920986023&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84920986023&partnerID=8YFLogxK

M3 - Article

C2 - 25519026

AN - SCOPUS:84920986023

VL - 68

SP - 323

EP - 329

JO - Acta Medica Okayama

JF - Acta Medica Okayama

SN - 0386-300X

IS - 6

ER -