Hemolysis complicating coil occlusion of patent ductus arteriosus

Hideshi Tomita, Shigeto Fuse, Teiji Akagi, Yasutoshi Matsumoto, Yasuo Murakami, Hisayuki Shiraya, Kazuyuki Koike, Masahiro Kamada, Tetsuro Kamiya, Kazuo Momma, Akira Ishizawa, Shunzo Chiba

Research output: Contribution to journalArticle

32 Citations (Scopus)

Abstract

We report on 5 patients who developed hemolysis (the Hemolysis group) following coil occlusion for PDA, and compare their data to 66 cases which were not complicated by hemolysis despite residual leak (the No Hemolysis group). A significant leak with a heart murmur was more frequent in the Hemolysis group than In the No Hemolysis group. The ratio of the sum of the loop diameter of coils to the minimal diameter of the ductus (C/D) in those who developed persistent hemolysis that needed a second intervention (2.2 ± 0.4) was significantly smaller than in the No Hemolysis group (3.1 ± 1.1). Persistent hemolysis can occur if a significant residual shunt remains after implantation of coils with small C/D. As this complication may be avoided by complete closure or, if not complete, a minimal leak, we should be careful to make residual leaks as small as possible by the use of multiple coils.

Original languageEnglish
Pages (from-to)50-54
Number of pages5
JournalCatheterization and Cardiovascular Diagnosis
Volume43
Issue number1
DOIs
Publication statusPublished - Jan 1998
Externally publishedYes

Fingerprint

Patent Ductus Arteriosus
Hemolysis
Heart Murmurs

Keywords

  • Coil occlusion
  • Complication
  • Hemolysis
  • Patent ductus arteriosus

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Hemolysis complicating coil occlusion of patent ductus arteriosus. / Tomita, Hideshi; Fuse, Shigeto; Akagi, Teiji; Matsumoto, Yasutoshi; Murakami, Yasuo; Shiraya, Hisayuki; Koike, Kazuyuki; Kamada, Masahiro; Kamiya, Tetsuro; Momma, Kazuo; Ishizawa, Akira; Chiba, Shunzo.

In: Catheterization and Cardiovascular Diagnosis, Vol. 43, No. 1, 01.1998, p. 50-54.

Research output: Contribution to journalArticle

Tomita, H, Fuse, S, Akagi, T, Matsumoto, Y, Murakami, Y, Shiraya, H, Koike, K, Kamada, M, Kamiya, T, Momma, K, Ishizawa, A & Chiba, S 1998, 'Hemolysis complicating coil occlusion of patent ductus arteriosus', Catheterization and Cardiovascular Diagnosis, vol. 43, no. 1, pp. 50-54. https://doi.org/10.1002/(SICI)1097-0304(199801)43:1<50::AID-CCD14>3.0.CO;2-X
Tomita, Hideshi ; Fuse, Shigeto ; Akagi, Teiji ; Matsumoto, Yasutoshi ; Murakami, Yasuo ; Shiraya, Hisayuki ; Koike, Kazuyuki ; Kamada, Masahiro ; Kamiya, Tetsuro ; Momma, Kazuo ; Ishizawa, Akira ; Chiba, Shunzo. / Hemolysis complicating coil occlusion of patent ductus arteriosus. In: Catheterization and Cardiovascular Diagnosis. 1998 ; Vol. 43, No. 1. pp. 50-54.
@article{ee407fa9a22445328531fb204efa6812,
title = "Hemolysis complicating coil occlusion of patent ductus arteriosus",
abstract = "We report on 5 patients who developed hemolysis (the Hemolysis group) following coil occlusion for PDA, and compare their data to 66 cases which were not complicated by hemolysis despite residual leak (the No Hemolysis group). A significant leak with a heart murmur was more frequent in the Hemolysis group than In the No Hemolysis group. The ratio of the sum of the loop diameter of coils to the minimal diameter of the ductus (C/D) in those who developed persistent hemolysis that needed a second intervention (2.2 ± 0.4) was significantly smaller than in the No Hemolysis group (3.1 ± 1.1). Persistent hemolysis can occur if a significant residual shunt remains after implantation of coils with small C/D. As this complication may be avoided by complete closure or, if not complete, a minimal leak, we should be careful to make residual leaks as small as possible by the use of multiple coils.",
keywords = "Coil occlusion, Complication, Hemolysis, Patent ductus arteriosus",
author = "Hideshi Tomita and Shigeto Fuse and Teiji Akagi and Yasutoshi Matsumoto and Yasuo Murakami and Hisayuki Shiraya and Kazuyuki Koike and Masahiro Kamada and Tetsuro Kamiya and Kazuo Momma and Akira Ishizawa and Shunzo Chiba",
year = "1998",
month = "1",
doi = "10.1002/(SICI)1097-0304(199801)43:1<50::AID-CCD14>3.0.CO;2-X",
language = "English",
volume = "43",
pages = "50--54",
journal = "Catheterization and Cardiovascular Interventions",
issn = "1522-1946",
publisher = "Wiley-Liss Inc.",
number = "1",

}

TY - JOUR

T1 - Hemolysis complicating coil occlusion of patent ductus arteriosus

AU - Tomita, Hideshi

AU - Fuse, Shigeto

AU - Akagi, Teiji

AU - Matsumoto, Yasutoshi

AU - Murakami, Yasuo

AU - Shiraya, Hisayuki

AU - Koike, Kazuyuki

AU - Kamada, Masahiro

AU - Kamiya, Tetsuro

AU - Momma, Kazuo

AU - Ishizawa, Akira

AU - Chiba, Shunzo

PY - 1998/1

Y1 - 1998/1

N2 - We report on 5 patients who developed hemolysis (the Hemolysis group) following coil occlusion for PDA, and compare their data to 66 cases which were not complicated by hemolysis despite residual leak (the No Hemolysis group). A significant leak with a heart murmur was more frequent in the Hemolysis group than In the No Hemolysis group. The ratio of the sum of the loop diameter of coils to the minimal diameter of the ductus (C/D) in those who developed persistent hemolysis that needed a second intervention (2.2 ± 0.4) was significantly smaller than in the No Hemolysis group (3.1 ± 1.1). Persistent hemolysis can occur if a significant residual shunt remains after implantation of coils with small C/D. As this complication may be avoided by complete closure or, if not complete, a minimal leak, we should be careful to make residual leaks as small as possible by the use of multiple coils.

AB - We report on 5 patients who developed hemolysis (the Hemolysis group) following coil occlusion for PDA, and compare their data to 66 cases which were not complicated by hemolysis despite residual leak (the No Hemolysis group). A significant leak with a heart murmur was more frequent in the Hemolysis group than In the No Hemolysis group. The ratio of the sum of the loop diameter of coils to the minimal diameter of the ductus (C/D) in those who developed persistent hemolysis that needed a second intervention (2.2 ± 0.4) was significantly smaller than in the No Hemolysis group (3.1 ± 1.1). Persistent hemolysis can occur if a significant residual shunt remains after implantation of coils with small C/D. As this complication may be avoided by complete closure or, if not complete, a minimal leak, we should be careful to make residual leaks as small as possible by the use of multiple coils.

KW - Coil occlusion

KW - Complication

KW - Hemolysis

KW - Patent ductus arteriosus

UR - http://www.scopus.com/inward/record.url?scp=0031962651&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031962651&partnerID=8YFLogxK

U2 - 10.1002/(SICI)1097-0304(199801)43:1<50::AID-CCD14>3.0.CO;2-X

DO - 10.1002/(SICI)1097-0304(199801)43:1<50::AID-CCD14>3.0.CO;2-X

M3 - Article

C2 - 9473189

AN - SCOPUS:0031962651

VL - 43

SP - 50

EP - 54

JO - Catheterization and Cardiovascular Interventions

JF - Catheterization and Cardiovascular Interventions

SN - 1522-1946

IS - 1

ER -