Half-value layer measurement: Simple process method using radiochromic film

T. Gotanda, T. Katsuda, R. Gotanda, A. Tabuchi, K. Yamamoto, T. Kuwano, H. Yatake, K. Kashiyama, Yoshihiro Takeda

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

Although the half-value layer (HVL) is one of the important parameters for QA and QC, constant monitoring has not been performed because the measurements using an ionization chamber (IC) are time-consuming and complicated. To solve these problems, the use of radiochromic film (GAFCHROMIC XR TYPE R: GAF-R) with step-shaped aluminum (Al) filters, referred to herein as the simple process method, has been developed. The measurement X-ray tube voltages were 120 kV, 100 kV, and 80 kV. The Al filter area, the full exposure area, and the unexposed area were set on the GAF-R so as to obtain correct data. The HVL was evaluated using the density attenuation ratio. The HVLs obtained using the GAF-R and an IC dosimeter were compared. HVLs with X-ray tube voltages of 120 kV, 100 kV, and 80 kV using the GAF-R were 4.10 mm, 3.55 mm and 2.97 mm, respectively. The difference ratios of the HVLs using the GAF-R and the IC were 1.2%, 7.6%, and 10.0%, respectively. The HVL at 120 kV can be routinely and quickly measured using the simple process method. Therefore, an IC dosimeter is not needed for HVL measurements for QA and QC However, the HVL measurements of low energy (100 kV and 80 kV) need attention.

Original languageEnglish
Pages (from-to)150-158
Number of pages9
JournalAustralasian Physical and Engineering Sciences in Medicine
Volume32
Issue number3
DOIs
Publication statusPublished - 2009

Fingerprint

Ionization chambers
Aluminum
X-Rays
X ray tubes
Dosimeters
Electric potential
Radiation Dosimeters
Monitoring

Keywords

  • Effective energy
  • Half-value layer
  • Quality assurance
  • Quality control
  • Radiochromic film

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Half-value layer measurement : Simple process method using radiochromic film. / Gotanda, T.; Katsuda, T.; Gotanda, R.; Tabuchi, A.; Yamamoto, K.; Kuwano, T.; Yatake, H.; Kashiyama, K.; Takeda, Yoshihiro.

In: Australasian Physical and Engineering Sciences in Medicine, Vol. 32, No. 3, 2009, p. 150-158.

Research output: Contribution to journalArticle

Gotanda, T, Katsuda, T, Gotanda, R, Tabuchi, A, Yamamoto, K, Kuwano, T, Yatake, H, Kashiyama, K & Takeda, Y 2009, 'Half-value layer measurement: Simple process method using radiochromic film', Australasian Physical and Engineering Sciences in Medicine, vol. 32, no. 3, pp. 150-158. https://doi.org/10.1007/BF03178642
Gotanda, T. ; Katsuda, T. ; Gotanda, R. ; Tabuchi, A. ; Yamamoto, K. ; Kuwano, T. ; Yatake, H. ; Kashiyama, K. ; Takeda, Yoshihiro. / Half-value layer measurement : Simple process method using radiochromic film. In: Australasian Physical and Engineering Sciences in Medicine. 2009 ; Vol. 32, No. 3. pp. 150-158.
@article{e91618d72230470093e31743d8de8c65,
title = "Half-value layer measurement: Simple process method using radiochromic film",
abstract = "Although the half-value layer (HVL) is one of the important parameters for QA and QC, constant monitoring has not been performed because the measurements using an ionization chamber (IC) are time-consuming and complicated. To solve these problems, the use of radiochromic film (GAFCHROMIC XR TYPE R: GAF-R) with step-shaped aluminum (Al) filters, referred to herein as the simple process method, has been developed. The measurement X-ray tube voltages were 120 kV, 100 kV, and 80 kV. The Al filter area, the full exposure area, and the unexposed area were set on the GAF-R so as to obtain correct data. The HVL was evaluated using the density attenuation ratio. The HVLs obtained using the GAF-R and an IC dosimeter were compared. HVLs with X-ray tube voltages of 120 kV, 100 kV, and 80 kV using the GAF-R were 4.10 mm, 3.55 mm and 2.97 mm, respectively. The difference ratios of the HVLs using the GAF-R and the IC were 1.2{\%}, 7.6{\%}, and 10.0{\%}, respectively. The HVL at 120 kV can be routinely and quickly measured using the simple process method. Therefore, an IC dosimeter is not needed for HVL measurements for QA and QC However, the HVL measurements of low energy (100 kV and 80 kV) need attention.",
keywords = "Effective energy, Half-value layer, Quality assurance, Quality control, Radiochromic film",
author = "T. Gotanda and T. Katsuda and R. Gotanda and A. Tabuchi and K. Yamamoto and T. Kuwano and H. Yatake and K. Kashiyama and Yoshihiro Takeda",
year = "2009",
doi = "10.1007/BF03178642",
language = "English",
volume = "32",
pages = "150--158",
journal = "Australasian Physical and Engineering Sciences in Medicine",
issn = "0158-9938",
publisher = "Springer Netherlands",
number = "3",

}

TY - JOUR

T1 - Half-value layer measurement

T2 - Simple process method using radiochromic film

AU - Gotanda, T.

AU - Katsuda, T.

AU - Gotanda, R.

AU - Tabuchi, A.

AU - Yamamoto, K.

AU - Kuwano, T.

AU - Yatake, H.

AU - Kashiyama, K.

AU - Takeda, Yoshihiro

PY - 2009

Y1 - 2009

N2 - Although the half-value layer (HVL) is one of the important parameters for QA and QC, constant monitoring has not been performed because the measurements using an ionization chamber (IC) are time-consuming and complicated. To solve these problems, the use of radiochromic film (GAFCHROMIC XR TYPE R: GAF-R) with step-shaped aluminum (Al) filters, referred to herein as the simple process method, has been developed. The measurement X-ray tube voltages were 120 kV, 100 kV, and 80 kV. The Al filter area, the full exposure area, and the unexposed area were set on the GAF-R so as to obtain correct data. The HVL was evaluated using the density attenuation ratio. The HVLs obtained using the GAF-R and an IC dosimeter were compared. HVLs with X-ray tube voltages of 120 kV, 100 kV, and 80 kV using the GAF-R were 4.10 mm, 3.55 mm and 2.97 mm, respectively. The difference ratios of the HVLs using the GAF-R and the IC were 1.2%, 7.6%, and 10.0%, respectively. The HVL at 120 kV can be routinely and quickly measured using the simple process method. Therefore, an IC dosimeter is not needed for HVL measurements for QA and QC However, the HVL measurements of low energy (100 kV and 80 kV) need attention.

AB - Although the half-value layer (HVL) is one of the important parameters for QA and QC, constant monitoring has not been performed because the measurements using an ionization chamber (IC) are time-consuming and complicated. To solve these problems, the use of radiochromic film (GAFCHROMIC XR TYPE R: GAF-R) with step-shaped aluminum (Al) filters, referred to herein as the simple process method, has been developed. The measurement X-ray tube voltages were 120 kV, 100 kV, and 80 kV. The Al filter area, the full exposure area, and the unexposed area were set on the GAF-R so as to obtain correct data. The HVL was evaluated using the density attenuation ratio. The HVLs obtained using the GAF-R and an IC dosimeter were compared. HVLs with X-ray tube voltages of 120 kV, 100 kV, and 80 kV using the GAF-R were 4.10 mm, 3.55 mm and 2.97 mm, respectively. The difference ratios of the HVLs using the GAF-R and the IC were 1.2%, 7.6%, and 10.0%, respectively. The HVL at 120 kV can be routinely and quickly measured using the simple process method. Therefore, an IC dosimeter is not needed for HVL measurements for QA and QC However, the HVL measurements of low energy (100 kV and 80 kV) need attention.

KW - Effective energy

KW - Half-value layer

KW - Quality assurance

KW - Quality control

KW - Radiochromic film

UR - http://www.scopus.com/inward/record.url?scp=71049116086&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=71049116086&partnerID=8YFLogxK

U2 - 10.1007/BF03178642

DO - 10.1007/BF03178642

M3 - Article

C2 - 19873939

AN - SCOPUS:71049116086

VL - 32

SP - 150

EP - 158

JO - Australasian Physical and Engineering Sciences in Medicine

JF - Australasian Physical and Engineering Sciences in Medicine

SN - 0158-9938

IS - 3

ER -