Development of MRI phantom equivalent to human tissues for 3.0-T MRI

Kengo Hattori, Yusuke Ikemoto, Wataru Takao, Seiichiro Ohno, Takashi Harimoto, Susumu Kanazawa, Masataka Oita, Koichi Shibuya, Masahiro Kuroda, Hirokazu Kato

Research output: Contribution to journalArticle

25 Citations (Scopus)

Abstract

Purpose: A 3.0-T MRI phantom (called the CAGN-3.0T phantom) having human-equivalent relaxation times and human-equivalent conductivity was developed. Methods: The ingredients of the phantom are carrageenan (as a gelatinizer), agarose (as a T2-relaxation modifier), GdCl3 (as a T1-relaxation modifier), NaCl (as a conductivity modifier), and NaN3 (as an antiseptic). Numerous samples with varying concentrations of agarose, GdCl3, and NaCl were prepared, and T 1 and T2 values were measured using 3.0-T MRI. Results: The T1 values of the CAGN-3.0T phantom were unaffected by NaCl, while the T2 values were only slightly affected. Based on the measured data, empirical formulae were devised to express the relationships between the concentrations of agarose, GdCl3, and NaCl and the relaxation times. The formula for expressing the conductivity of the CAGN-3.0T phantom was obtained. Conclusions: By adjustments to the concentrations of agarose, GdCl3, and NaCl, the relaxation times and conductivity of almost all types of human tissues can be simulated by CAGN-3.0T phantoms. The phantoms have T1 values of 395-2601 ms, T2 values of 29-334 ms, and conductivity of 0.27-1.26 S/m when concentrations of agarose, GdCl3, and NaCl are varied from 0 to 2.0 w/w%, 0 to 180 μmol/kg, and 0 to 0.7 w/w%, respectively. The CAGN-3.0T phantom has sufficient strength to replicate the torso without using reinforcing agents, and can be cut with a knife into any shape.

Original languageEnglish
Article number032303
JournalMedical Physics
Volume40
Issue number3
DOIs
Publication statusPublished - Mar 2013

Fingerprint

Sepharose
Torso
Sodium Azide
Local Anti-Infective Agents
Carrageenan

Keywords

  • 3.0T MRI
  • conductivity
  • phantom
  • T
  • T

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Hattori, K., Ikemoto, Y., Takao, W., Ohno, S., Harimoto, T., Kanazawa, S., ... Kato, H. (2013). Development of MRI phantom equivalent to human tissues for 3.0-T MRI. Medical Physics, 40(3), [032303]. https://doi.org/10.1118/1.4790023

Development of MRI phantom equivalent to human tissues for 3.0-T MRI. / Hattori, Kengo; Ikemoto, Yusuke; Takao, Wataru; Ohno, Seiichiro; Harimoto, Takashi; Kanazawa, Susumu; Oita, Masataka; Shibuya, Koichi; Kuroda, Masahiro; Kato, Hirokazu.

In: Medical Physics, Vol. 40, No. 3, 032303, 03.2013.

Research output: Contribution to journalArticle

Hattori, Kengo ; Ikemoto, Yusuke ; Takao, Wataru ; Ohno, Seiichiro ; Harimoto, Takashi ; Kanazawa, Susumu ; Oita, Masataka ; Shibuya, Koichi ; Kuroda, Masahiro ; Kato, Hirokazu. / Development of MRI phantom equivalent to human tissues for 3.0-T MRI. In: Medical Physics. 2013 ; Vol. 40, No. 3.
@article{9906344e1dd94df6a29e2ec01282e106,
title = "Development of MRI phantom equivalent to human tissues for 3.0-T MRI",
abstract = "Purpose: A 3.0-T MRI phantom (called the CAGN-3.0T phantom) having human-equivalent relaxation times and human-equivalent conductivity was developed. Methods: The ingredients of the phantom are carrageenan (as a gelatinizer), agarose (as a T2-relaxation modifier), GdCl3 (as a T1-relaxation modifier), NaCl (as a conductivity modifier), and NaN3 (as an antiseptic). Numerous samples with varying concentrations of agarose, GdCl3, and NaCl were prepared, and T 1 and T2 values were measured using 3.0-T MRI. Results: The T1 values of the CAGN-3.0T phantom were unaffected by NaCl, while the T2 values were only slightly affected. Based on the measured data, empirical formulae were devised to express the relationships between the concentrations of agarose, GdCl3, and NaCl and the relaxation times. The formula for expressing the conductivity of the CAGN-3.0T phantom was obtained. Conclusions: By adjustments to the concentrations of agarose, GdCl3, and NaCl, the relaxation times and conductivity of almost all types of human tissues can be simulated by CAGN-3.0T phantoms. The phantoms have T1 values of 395-2601 ms, T2 values of 29-334 ms, and conductivity of 0.27-1.26 S/m when concentrations of agarose, GdCl3, and NaCl are varied from 0 to 2.0 w/w{\%}, 0 to 180 μmol/kg, and 0 to 0.7 w/w{\%}, respectively. The CAGN-3.0T phantom has sufficient strength to replicate the torso without using reinforcing agents, and can be cut with a knife into any shape.",
keywords = "3.0T MRI, conductivity, phantom, T, T",
author = "Kengo Hattori and Yusuke Ikemoto and Wataru Takao and Seiichiro Ohno and Takashi Harimoto and Susumu Kanazawa and Masataka Oita and Koichi Shibuya and Masahiro Kuroda and Hirokazu Kato",
year = "2013",
month = "3",
doi = "10.1118/1.4790023",
language = "English",
volume = "40",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "3",

}

TY - JOUR

T1 - Development of MRI phantom equivalent to human tissues for 3.0-T MRI

AU - Hattori, Kengo

AU - Ikemoto, Yusuke

AU - Takao, Wataru

AU - Ohno, Seiichiro

AU - Harimoto, Takashi

AU - Kanazawa, Susumu

AU - Oita, Masataka

AU - Shibuya, Koichi

AU - Kuroda, Masahiro

AU - Kato, Hirokazu

PY - 2013/3

Y1 - 2013/3

N2 - Purpose: A 3.0-T MRI phantom (called the CAGN-3.0T phantom) having human-equivalent relaxation times and human-equivalent conductivity was developed. Methods: The ingredients of the phantom are carrageenan (as a gelatinizer), agarose (as a T2-relaxation modifier), GdCl3 (as a T1-relaxation modifier), NaCl (as a conductivity modifier), and NaN3 (as an antiseptic). Numerous samples with varying concentrations of agarose, GdCl3, and NaCl were prepared, and T 1 and T2 values were measured using 3.0-T MRI. Results: The T1 values of the CAGN-3.0T phantom were unaffected by NaCl, while the T2 values were only slightly affected. Based on the measured data, empirical formulae were devised to express the relationships between the concentrations of agarose, GdCl3, and NaCl and the relaxation times. The formula for expressing the conductivity of the CAGN-3.0T phantom was obtained. Conclusions: By adjustments to the concentrations of agarose, GdCl3, and NaCl, the relaxation times and conductivity of almost all types of human tissues can be simulated by CAGN-3.0T phantoms. The phantoms have T1 values of 395-2601 ms, T2 values of 29-334 ms, and conductivity of 0.27-1.26 S/m when concentrations of agarose, GdCl3, and NaCl are varied from 0 to 2.0 w/w%, 0 to 180 μmol/kg, and 0 to 0.7 w/w%, respectively. The CAGN-3.0T phantom has sufficient strength to replicate the torso without using reinforcing agents, and can be cut with a knife into any shape.

AB - Purpose: A 3.0-T MRI phantom (called the CAGN-3.0T phantom) having human-equivalent relaxation times and human-equivalent conductivity was developed. Methods: The ingredients of the phantom are carrageenan (as a gelatinizer), agarose (as a T2-relaxation modifier), GdCl3 (as a T1-relaxation modifier), NaCl (as a conductivity modifier), and NaN3 (as an antiseptic). Numerous samples with varying concentrations of agarose, GdCl3, and NaCl were prepared, and T 1 and T2 values were measured using 3.0-T MRI. Results: The T1 values of the CAGN-3.0T phantom were unaffected by NaCl, while the T2 values were only slightly affected. Based on the measured data, empirical formulae were devised to express the relationships between the concentrations of agarose, GdCl3, and NaCl and the relaxation times. The formula for expressing the conductivity of the CAGN-3.0T phantom was obtained. Conclusions: By adjustments to the concentrations of agarose, GdCl3, and NaCl, the relaxation times and conductivity of almost all types of human tissues can be simulated by CAGN-3.0T phantoms. The phantoms have T1 values of 395-2601 ms, T2 values of 29-334 ms, and conductivity of 0.27-1.26 S/m when concentrations of agarose, GdCl3, and NaCl are varied from 0 to 2.0 w/w%, 0 to 180 μmol/kg, and 0 to 0.7 w/w%, respectively. The CAGN-3.0T phantom has sufficient strength to replicate the torso without using reinforcing agents, and can be cut with a knife into any shape.

KW - 3.0T MRI

KW - conductivity

KW - phantom

KW - T

KW - T

UR - http://www.scopus.com/inward/record.url?scp=84874768123&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84874768123&partnerID=8YFLogxK

U2 - 10.1118/1.4790023

DO - 10.1118/1.4790023

M3 - Article

C2 - 23464335

AN - SCOPUS:84874768123

VL - 40

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 3

M1 - 032303

ER -