Determination of bisphenol A in effluents of hemodialyzers

Hiroko Yamasaki, Yoshio Nagake, Hirofumi Makino

Research output: Contribution to journalArticle

33 Citations (Scopus)

Abstract

This study was designed to determine whether bisphenol A (BPA) is eluted from hemodialyzers in which polycarbonate and polysulfone based on BPA are used as materials. Four types of polysulfone hemodialyzer (PS hemodialyzer: PS-A, PS-B, PS-C and PS-D) and an ethylene-vinyl alcohol copolymer dialyzer (EVAL hemodialyzer) were used in this study. In the PS-C, PS-D and EVAL hemodialyzers, polycarbonate was used in the case headers at both ends of the hemodialyzer. In in vitro experiments, the hemodialyzers were filled with reverse osmotic water, and BPA concentrations were measured. Saline solution (200 ml) was then circulated through a blood circuit tube connected to the hemodialyzer, and BPA concentrations in the saline solution were measured. In in vivo experiments, BPA concentrations in whole blood samples from hemodialysis patients treated with PS-C (n = 3) and PS-D (n = 3) hemodialyzers were measured. In in vitro experiments, BPA was detected in the effluents of the PS-C, PS-D and EVAL hemodialyzers. In in vivo experiments, BPA was detected in whole blood samples from hemodialysis patients treated with the PS-D hemodialyzer (mean value, 0.77 ppb). This is the first report of BPA elution from hemodialyzers in which polycarbonate and polysulfone are used, and also the first report of detection of BPA in whole blood samples from patients on hemodialysis.

Original languageEnglish
Pages (from-to)376-378
Number of pages3
JournalNephron
Volume88
Issue number4
DOIs
Publication statusPublished - Aug 11 2001

Fingerprint

Artificial Kidneys
polycarbonate
Renal Dialysis
Sodium Chloride
bisphenol A

Keywords

  • Bisphenol A
  • Endocrine disrupter
  • Hemodialysis

ASJC Scopus subject areas

  • Nephrology

Cite this

Determination of bisphenol A in effluents of hemodialyzers. / Yamasaki, Hiroko; Nagake, Yoshio; Makino, Hirofumi.

In: Nephron, Vol. 88, No. 4, 11.08.2001, p. 376-378.

Research output: Contribution to journalArticle

Yamasaki, Hiroko ; Nagake, Yoshio ; Makino, Hirofumi. / Determination of bisphenol A in effluents of hemodialyzers. In: Nephron. 2001 ; Vol. 88, No. 4. pp. 376-378.
@article{25ab893ce3bd4056a7cddba200e611e7,
title = "Determination of bisphenol A in effluents of hemodialyzers",
abstract = "This study was designed to determine whether bisphenol A (BPA) is eluted from hemodialyzers in which polycarbonate and polysulfone based on BPA are used as materials. Four types of polysulfone hemodialyzer (PS hemodialyzer: PS-A, PS-B, PS-C and PS-D) and an ethylene-vinyl alcohol copolymer dialyzer (EVAL hemodialyzer) were used in this study. In the PS-C, PS-D and EVAL hemodialyzers, polycarbonate was used in the case headers at both ends of the hemodialyzer. In in vitro experiments, the hemodialyzers were filled with reverse osmotic water, and BPA concentrations were measured. Saline solution (200 ml) was then circulated through a blood circuit tube connected to the hemodialyzer, and BPA concentrations in the saline solution were measured. In in vivo experiments, BPA concentrations in whole blood samples from hemodialysis patients treated with PS-C (n = 3) and PS-D (n = 3) hemodialyzers were measured. In in vitro experiments, BPA was detected in the effluents of the PS-C, PS-D and EVAL hemodialyzers. In in vivo experiments, BPA was detected in whole blood samples from hemodialysis patients treated with the PS-D hemodialyzer (mean value, 0.77 ppb). This is the first report of BPA elution from hemodialyzers in which polycarbonate and polysulfone are used, and also the first report of detection of BPA in whole blood samples from patients on hemodialysis.",
keywords = "Bisphenol A, Endocrine disrupter, Hemodialysis",
author = "Hiroko Yamasaki and Yoshio Nagake and Hirofumi Makino",
year = "2001",
month = "8",
day = "11",
doi = "10.1159/000046023",
language = "English",
volume = "88",
pages = "376--378",
journal = "Experimental Nephrology",
issn = "0028-2766",
publisher = "S. Karger AG",
number = "4",

}

TY - JOUR

T1 - Determination of bisphenol A in effluents of hemodialyzers

AU - Yamasaki, Hiroko

AU - Nagake, Yoshio

AU - Makino, Hirofumi

PY - 2001/8/11

Y1 - 2001/8/11

N2 - This study was designed to determine whether bisphenol A (BPA) is eluted from hemodialyzers in which polycarbonate and polysulfone based on BPA are used as materials. Four types of polysulfone hemodialyzer (PS hemodialyzer: PS-A, PS-B, PS-C and PS-D) and an ethylene-vinyl alcohol copolymer dialyzer (EVAL hemodialyzer) were used in this study. In the PS-C, PS-D and EVAL hemodialyzers, polycarbonate was used in the case headers at both ends of the hemodialyzer. In in vitro experiments, the hemodialyzers were filled with reverse osmotic water, and BPA concentrations were measured. Saline solution (200 ml) was then circulated through a blood circuit tube connected to the hemodialyzer, and BPA concentrations in the saline solution were measured. In in vivo experiments, BPA concentrations in whole blood samples from hemodialysis patients treated with PS-C (n = 3) and PS-D (n = 3) hemodialyzers were measured. In in vitro experiments, BPA was detected in the effluents of the PS-C, PS-D and EVAL hemodialyzers. In in vivo experiments, BPA was detected in whole blood samples from hemodialysis patients treated with the PS-D hemodialyzer (mean value, 0.77 ppb). This is the first report of BPA elution from hemodialyzers in which polycarbonate and polysulfone are used, and also the first report of detection of BPA in whole blood samples from patients on hemodialysis.

AB - This study was designed to determine whether bisphenol A (BPA) is eluted from hemodialyzers in which polycarbonate and polysulfone based on BPA are used as materials. Four types of polysulfone hemodialyzer (PS hemodialyzer: PS-A, PS-B, PS-C and PS-D) and an ethylene-vinyl alcohol copolymer dialyzer (EVAL hemodialyzer) were used in this study. In the PS-C, PS-D and EVAL hemodialyzers, polycarbonate was used in the case headers at both ends of the hemodialyzer. In in vitro experiments, the hemodialyzers were filled with reverse osmotic water, and BPA concentrations were measured. Saline solution (200 ml) was then circulated through a blood circuit tube connected to the hemodialyzer, and BPA concentrations in the saline solution were measured. In in vivo experiments, BPA concentrations in whole blood samples from hemodialysis patients treated with PS-C (n = 3) and PS-D (n = 3) hemodialyzers were measured. In in vitro experiments, BPA was detected in the effluents of the PS-C, PS-D and EVAL hemodialyzers. In in vivo experiments, BPA was detected in whole blood samples from hemodialysis patients treated with the PS-D hemodialyzer (mean value, 0.77 ppb). This is the first report of BPA elution from hemodialyzers in which polycarbonate and polysulfone are used, and also the first report of detection of BPA in whole blood samples from patients on hemodialysis.

KW - Bisphenol A

KW - Endocrine disrupter

KW - Hemodialysis

UR - http://www.scopus.com/inward/record.url?scp=0034929188&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034929188&partnerID=8YFLogxK

U2 - 10.1159/000046023

DO - 10.1159/000046023

M3 - Article

C2 - 11474234

AN - SCOPUS:0034929188

VL - 88

SP - 376

EP - 378

JO - Experimental Nephrology

JF - Experimental Nephrology

SN - 0028-2766

IS - 4