Characteristics of a Pneumatic bellows actuator for Colonoscopy

Shuichi Wakimoto, Hidehiro Kametani

Research output: Contribution to journalArticle

Abstract

Since colonoscopic insertion requires a high level of skill on the part of the examining physician, the patient may experience pain if the physician is not so highly skilled. In this study, our objective is to realize an easy-to-execute endoscopic insertion that does not rely on the physician’s skill. We develop a pneumatic actuator that gives the endoscope self-propulsion capability. The actuator is configured with three bellows made of polyethylene film. By applying pneumatic pressure to the bellows in a specific sequence to extend and contract them, the end of the actuator moves elliptically. Thus, the endoscope can be given propulsion performance by attaching several actuators and driving them in different phases. In this paper, the basic characteristics of the actuator are determined geometrically from the characteristics of the bellows, and then those characteristics are compared to the results of experiments. In addition, experiments are conducted in which a rod, used as a dummy endoscope, is transported through a pipe and inserted in an intestinal tractmodel. The experiments confirmand validate the proposed concept.

Original languageEnglish
Pages (from-to)479-486
Number of pages8
JournalInternational Journal of Automation Technology
Volume10
Issue number4
Publication statusPublished - 2016

Fingerprint

Bellows
Pneumatics
Endoscopy
Actuators
Propulsion
Pneumatic actuators
Experiments
Polyethylenes
Pipe

Keywords

  • Colonoscope insertion assist device
  • Film actuator
  • Pneumatic actuator

ASJC Scopus subject areas

  • Mechanical Engineering
  • Industrial and Manufacturing Engineering

Cite this

Characteristics of a Pneumatic bellows actuator for Colonoscopy. / Wakimoto, Shuichi; Kametani, Hidehiro.

In: International Journal of Automation Technology, Vol. 10, No. 4, 2016, p. 479-486.

Research output: Contribution to journalArticle

@article{82b5851f6fc1442dbbcc031121166cdb,
title = "Characteristics of a Pneumatic bellows actuator for Colonoscopy",
abstract = "Since colonoscopic insertion requires a high level of skill on the part of the examining physician, the patient may experience pain if the physician is not so highly skilled. In this study, our objective is to realize an easy-to-execute endoscopic insertion that does not rely on the physician’s skill. We develop a pneumatic actuator that gives the endoscope self-propulsion capability. The actuator is configured with three bellows made of polyethylene film. By applying pneumatic pressure to the bellows in a specific sequence to extend and contract them, the end of the actuator moves elliptically. Thus, the endoscope can be given propulsion performance by attaching several actuators and driving them in different phases. In this paper, the basic characteristics of the actuator are determined geometrically from the characteristics of the bellows, and then those characteristics are compared to the results of experiments. In addition, experiments are conducted in which a rod, used as a dummy endoscope, is transported through a pipe and inserted in an intestinal tractmodel. The experiments confirmand validate the proposed concept.",
keywords = "Colonoscope insertion assist device, Film actuator, Pneumatic actuator",
author = "Shuichi Wakimoto and Hidehiro Kametani",
year = "2016",
language = "English",
volume = "10",
pages = "479--486",
journal = "International Journal of Automation Technology",
issn = "1881-7629",
publisher = "Fuji Technology Press",
number = "4",

}

TY - JOUR

T1 - Characteristics of a Pneumatic bellows actuator for Colonoscopy

AU - Wakimoto, Shuichi

AU - Kametani, Hidehiro

PY - 2016

Y1 - 2016

N2 - Since colonoscopic insertion requires a high level of skill on the part of the examining physician, the patient may experience pain if the physician is not so highly skilled. In this study, our objective is to realize an easy-to-execute endoscopic insertion that does not rely on the physician’s skill. We develop a pneumatic actuator that gives the endoscope self-propulsion capability. The actuator is configured with three bellows made of polyethylene film. By applying pneumatic pressure to the bellows in a specific sequence to extend and contract them, the end of the actuator moves elliptically. Thus, the endoscope can be given propulsion performance by attaching several actuators and driving them in different phases. In this paper, the basic characteristics of the actuator are determined geometrically from the characteristics of the bellows, and then those characteristics are compared to the results of experiments. In addition, experiments are conducted in which a rod, used as a dummy endoscope, is transported through a pipe and inserted in an intestinal tractmodel. The experiments confirmand validate the proposed concept.

AB - Since colonoscopic insertion requires a high level of skill on the part of the examining physician, the patient may experience pain if the physician is not so highly skilled. In this study, our objective is to realize an easy-to-execute endoscopic insertion that does not rely on the physician’s skill. We develop a pneumatic actuator that gives the endoscope self-propulsion capability. The actuator is configured with three bellows made of polyethylene film. By applying pneumatic pressure to the bellows in a specific sequence to extend and contract them, the end of the actuator moves elliptically. Thus, the endoscope can be given propulsion performance by attaching several actuators and driving them in different phases. In this paper, the basic characteristics of the actuator are determined geometrically from the characteristics of the bellows, and then those characteristics are compared to the results of experiments. In addition, experiments are conducted in which a rod, used as a dummy endoscope, is transported through a pipe and inserted in an intestinal tractmodel. The experiments confirmand validate the proposed concept.

KW - Colonoscope insertion assist device

KW - Film actuator

KW - Pneumatic actuator

UR - http://www.scopus.com/inward/record.url?scp=84978115919&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84978115919&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:84978115919

VL - 10

SP - 479

EP - 486

JO - International Journal of Automation Technology

JF - International Journal of Automation Technology

SN - 1881-7629

IS - 4

ER -