Axial length measurement using partial coherence interferometry in myopic children: Repeatability of the measurement and comparison with refractive components

Shuhei Kimura, Satoshi Hasebe, Manabu Miyata, Ichiro Hamasaki, Hiroshi Ohtsuki

Research output: Contribution to journalArticle

19 Citations (Scopus)

Abstract

Purpose: To evaluate the test-retest repeatability of axial length (AL) measurements made with the IOLMaster in children with a wide range of myopic refractive errors, and to clarify the relationship between the AL and other refractive components. Methods: The subjects were 95 children (mean age ± SD, 10.8 ± 1.3 years; range, 7-13 years; mean refractive error; -4.37 ± 1.43 D; range, -1.50 to -8.19 D) who participated in a myopia-control trial. The AL of the right eye was measured three times using an IOLMaster, and the mean value was regarded as the representative measurement. After 5 min, the measurement was performed again, and the repeatability was evaluated by analyzing the distribution of differences between the two measurements. The relationships between the AL and age, cycloplegic autorefraction (RE), and corneal radius of curvature (CR) were also examined. Results: The repeatability of the IOLMaster measurements was ±0.05 mm (corresponding to a refractive error of ±0.12 D) and was not affected by age or RE. AL was negatively correlated with RE (RE = -0.68 x AL + 12.74, r = -0.37) and positively correlated with CR (CR = 0.21 x AL + 2.53, r = 0.69). The highest correlation was found between the ratio of AL to CR and RE (AL/CR = -0.04 x RE + 3.08, r = -0.76). No association was observed between age and AL, nor between CR and RE. Conclusions: The IOLMaster provides high repeatability in AL measurement in myopic children. The associations between AL and each refractive component found in this study were consistent with those in previous studies that used the ultrasound A-mode method, except for slightly higher AL/CR ratios in our study.

Original languageEnglish
Pages (from-to)105-110
Number of pages6
JournalJapanese Journal of Ophthalmology
Volume51
Issue number2
DOIs
Publication statusPublished - Mar 2007

Fingerprint

Interferometry
Refractive Errors
Eye Axial Length
Mydriatics
Myopia

Keywords

  • Axial length
  • Myopic children
  • Partial coherence interferometry
  • Refractive components
  • Repeatability

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Axial length measurement using partial coherence interferometry in myopic children : Repeatability of the measurement and comparison with refractive components. / Kimura, Shuhei; Hasebe, Satoshi; Miyata, Manabu; Hamasaki, Ichiro; Ohtsuki, Hiroshi.

In: Japanese Journal of Ophthalmology, Vol. 51, No. 2, 03.2007, p. 105-110.

Research output: Contribution to journalArticle

@article{426ca7b25e5247c08eaa06d5f1e1294d,
title = "Axial length measurement using partial coherence interferometry in myopic children: Repeatability of the measurement and comparison with refractive components",
abstract = "Purpose: To evaluate the test-retest repeatability of axial length (AL) measurements made with the IOLMaster in children with a wide range of myopic refractive errors, and to clarify the relationship between the AL and other refractive components. Methods: The subjects were 95 children (mean age ± SD, 10.8 ± 1.3 years; range, 7-13 years; mean refractive error; -4.37 ± 1.43 D; range, -1.50 to -8.19 D) who participated in a myopia-control trial. The AL of the right eye was measured three times using an IOLMaster, and the mean value was regarded as the representative measurement. After 5 min, the measurement was performed again, and the repeatability was evaluated by analyzing the distribution of differences between the two measurements. The relationships between the AL and age, cycloplegic autorefraction (RE), and corneal radius of curvature (CR) were also examined. Results: The repeatability of the IOLMaster measurements was ±0.05 mm (corresponding to a refractive error of ±0.12 D) and was not affected by age or RE. AL was negatively correlated with RE (RE = -0.68 x AL + 12.74, r = -0.37) and positively correlated with CR (CR = 0.21 x AL + 2.53, r = 0.69). The highest correlation was found between the ratio of AL to CR and RE (AL/CR = -0.04 x RE + 3.08, r = -0.76). No association was observed between age and AL, nor between CR and RE. Conclusions: The IOLMaster provides high repeatability in AL measurement in myopic children. The associations between AL and each refractive component found in this study were consistent with those in previous studies that used the ultrasound A-mode method, except for slightly higher AL/CR ratios in our study.",
keywords = "Axial length, Myopic children, Partial coherence interferometry, Refractive components, Repeatability",
author = "Shuhei Kimura and Satoshi Hasebe and Manabu Miyata and Ichiro Hamasaki and Hiroshi Ohtsuki",
year = "2007",
month = "3",
doi = "10.1007/s10384-006-0410-5",
language = "English",
volume = "51",
pages = "105--110",
journal = "Japanese Journal of Ophthalmology",
issn = "0021-5155",
publisher = "Springer Japan",
number = "2",

}

TY - JOUR

T1 - Axial length measurement using partial coherence interferometry in myopic children

T2 - Repeatability of the measurement and comparison with refractive components

AU - Kimura, Shuhei

AU - Hasebe, Satoshi

AU - Miyata, Manabu

AU - Hamasaki, Ichiro

AU - Ohtsuki, Hiroshi

PY - 2007/3

Y1 - 2007/3

N2 - Purpose: To evaluate the test-retest repeatability of axial length (AL) measurements made with the IOLMaster in children with a wide range of myopic refractive errors, and to clarify the relationship between the AL and other refractive components. Methods: The subjects were 95 children (mean age ± SD, 10.8 ± 1.3 years; range, 7-13 years; mean refractive error; -4.37 ± 1.43 D; range, -1.50 to -8.19 D) who participated in a myopia-control trial. The AL of the right eye was measured three times using an IOLMaster, and the mean value was regarded as the representative measurement. After 5 min, the measurement was performed again, and the repeatability was evaluated by analyzing the distribution of differences between the two measurements. The relationships between the AL and age, cycloplegic autorefraction (RE), and corneal radius of curvature (CR) were also examined. Results: The repeatability of the IOLMaster measurements was ±0.05 mm (corresponding to a refractive error of ±0.12 D) and was not affected by age or RE. AL was negatively correlated with RE (RE = -0.68 x AL + 12.74, r = -0.37) and positively correlated with CR (CR = 0.21 x AL + 2.53, r = 0.69). The highest correlation was found between the ratio of AL to CR and RE (AL/CR = -0.04 x RE + 3.08, r = -0.76). No association was observed between age and AL, nor between CR and RE. Conclusions: The IOLMaster provides high repeatability in AL measurement in myopic children. The associations between AL and each refractive component found in this study were consistent with those in previous studies that used the ultrasound A-mode method, except for slightly higher AL/CR ratios in our study.

AB - Purpose: To evaluate the test-retest repeatability of axial length (AL) measurements made with the IOLMaster in children with a wide range of myopic refractive errors, and to clarify the relationship between the AL and other refractive components. Methods: The subjects were 95 children (mean age ± SD, 10.8 ± 1.3 years; range, 7-13 years; mean refractive error; -4.37 ± 1.43 D; range, -1.50 to -8.19 D) who participated in a myopia-control trial. The AL of the right eye was measured three times using an IOLMaster, and the mean value was regarded as the representative measurement. After 5 min, the measurement was performed again, and the repeatability was evaluated by analyzing the distribution of differences between the two measurements. The relationships between the AL and age, cycloplegic autorefraction (RE), and corneal radius of curvature (CR) were also examined. Results: The repeatability of the IOLMaster measurements was ±0.05 mm (corresponding to a refractive error of ±0.12 D) and was not affected by age or RE. AL was negatively correlated with RE (RE = -0.68 x AL + 12.74, r = -0.37) and positively correlated with CR (CR = 0.21 x AL + 2.53, r = 0.69). The highest correlation was found between the ratio of AL to CR and RE (AL/CR = -0.04 x RE + 3.08, r = -0.76). No association was observed between age and AL, nor between CR and RE. Conclusions: The IOLMaster provides high repeatability in AL measurement in myopic children. The associations between AL and each refractive component found in this study were consistent with those in previous studies that used the ultrasound A-mode method, except for slightly higher AL/CR ratios in our study.

KW - Axial length

KW - Myopic children

KW - Partial coherence interferometry

KW - Refractive components

KW - Repeatability

UR - http://www.scopus.com/inward/record.url?scp=34047238184&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34047238184&partnerID=8YFLogxK

U2 - 10.1007/s10384-006-0410-5

DO - 10.1007/s10384-006-0410-5

M3 - Article

C2 - 17401619

AN - SCOPUS:34047238184

VL - 51

SP - 105

EP - 110

JO - Japanese Journal of Ophthalmology

JF - Japanese Journal of Ophthalmology

SN - 0021-5155

IS - 2

ER -