Automation of pH measurement using a flow-through type differential pH sensor system based on pH-FET

Satoshi Mohri, Michihiro Nakamura, Keiji Naruse

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

Automatic pH measurement system was developed by using a flow-through type differential pH-FET sensor. The system was operated by repeating 60 sec cycle consisting of 36 sec of washing period, 4 sec of moving period, and 20 sec of measuring period. In washing period, the differential pH-FET sensor was washed with baseline solution. In measuring period, about 50 μ1 of sample solution was withdrawn to the space around a measurement pH-FET, and the change of differential source potential was measured. Absolute values of error for nine sample solutions with weak buffering ability were less than 0.146 pH, and were 0.071 pH in average. From this experiment, it was suggested that high-speed and automated pH measurement will be possible by using pH-FET sensors even for sample solutions of weak buffer ability.

Original languageEnglish
JournalIEEJ Transactions on Sensors and Micromachines
Volume127
Issue number8
Publication statusPublished - 2007

Fingerprint

pH sensors
Field effect transistors
Automation
Washing
Sensors
Experiments

Keywords

  • Automation
  • Differential pH sensor
  • ISFET
  • pH

ASJC Scopus subject areas

  • Electrical and Electronic Engineering
  • Mechanical Engineering

Cite this

Automation of pH measurement using a flow-through type differential pH sensor system based on pH-FET. / Mohri, Satoshi; Nakamura, Michihiro; Naruse, Keiji.

In: IEEJ Transactions on Sensors and Micromachines, Vol. 127, No. 8, 2007.

Research output: Contribution to journalArticle

@article{4266afba951f4a57af782ddc409c96e9,
title = "Automation of pH measurement using a flow-through type differential pH sensor system based on pH-FET",
abstract = "Automatic pH measurement system was developed by using a flow-through type differential pH-FET sensor. The system was operated by repeating 60 sec cycle consisting of 36 sec of washing period, 4 sec of moving period, and 20 sec of measuring period. In washing period, the differential pH-FET sensor was washed with baseline solution. In measuring period, about 50 μ1 of sample solution was withdrawn to the space around a measurement pH-FET, and the change of differential source potential was measured. Absolute values of error for nine sample solutions with weak buffering ability were less than 0.146 pH, and were 0.071 pH in average. From this experiment, it was suggested that high-speed and automated pH measurement will be possible by using pH-FET sensors even for sample solutions of weak buffer ability.",
keywords = "Automation, Differential pH sensor, ISFET, pH",
author = "Satoshi Mohri and Michihiro Nakamura and Keiji Naruse",
year = "2007",
language = "English",
volume = "127",
journal = "IEEJ Transactions on Sensors and Micromachines",
issn = "1341-8939",
publisher = "The Institute of Electrical Engineers of Japan",
number = "8",

}

TY - JOUR

T1 - Automation of pH measurement using a flow-through type differential pH sensor system based on pH-FET

AU - Mohri, Satoshi

AU - Nakamura, Michihiro

AU - Naruse, Keiji

PY - 2007

Y1 - 2007

N2 - Automatic pH measurement system was developed by using a flow-through type differential pH-FET sensor. The system was operated by repeating 60 sec cycle consisting of 36 sec of washing period, 4 sec of moving period, and 20 sec of measuring period. In washing period, the differential pH-FET sensor was washed with baseline solution. In measuring period, about 50 μ1 of sample solution was withdrawn to the space around a measurement pH-FET, and the change of differential source potential was measured. Absolute values of error for nine sample solutions with weak buffering ability were less than 0.146 pH, and were 0.071 pH in average. From this experiment, it was suggested that high-speed and automated pH measurement will be possible by using pH-FET sensors even for sample solutions of weak buffer ability.

AB - Automatic pH measurement system was developed by using a flow-through type differential pH-FET sensor. The system was operated by repeating 60 sec cycle consisting of 36 sec of washing period, 4 sec of moving period, and 20 sec of measuring period. In washing period, the differential pH-FET sensor was washed with baseline solution. In measuring period, about 50 μ1 of sample solution was withdrawn to the space around a measurement pH-FET, and the change of differential source potential was measured. Absolute values of error for nine sample solutions with weak buffering ability were less than 0.146 pH, and were 0.071 pH in average. From this experiment, it was suggested that high-speed and automated pH measurement will be possible by using pH-FET sensors even for sample solutions of weak buffer ability.

KW - Automation

KW - Differential pH sensor

KW - ISFET

KW - pH

UR - http://www.scopus.com/inward/record.url?scp=34548292386&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34548292386&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:34548292386

VL - 127

JO - IEEJ Transactions on Sensors and Micromachines

JF - IEEJ Transactions on Sensors and Micromachines

SN - 1341-8939

IS - 8

ER -