Accuracy of absorbed dose calculation and measurement of scatter factors with different depth of mini-phantoms using a pinpoint ionization chamber

Masataka Oita, Yoshiharu Watanabe, Katsuhisa Fujita, Teruo Furuya, Junya Nankumo, Makoto Oda

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The output factor of high-energy X-ray machines varies with collimation. According to Khan's theory, collimator and phantom scatter factors contribute to total scatter factor. For precise X-ray irradiation, the two factors need to be taken into consideration. To obtain proper factors, we made two original polystyrene cylindrical mini-phantoms. These phantoms are both 4 cm in diameter and have a pinpoint ion chamber placed at a depth of 5 cm and 10 cm, respectively. Using a 6 MV X-ray machine, collimator scatter factors were calculated for various field arrangements (i.e., field sizes ranging from 4 cm x 4 cm to 40 cm x 40 cm at isocenter). To determine if calculated values were appropriate, we measured point doses of 20 X-ray irradiation patterns using a Farmer-type ion chamber with a water equivalent phantom at depths of 5 cm and 10 cm, respectively. Two hundred MUs were irradiated to the above-mentioned depths for each field. Based on the measured doses, variations were obtained for four calculation methods. Accounting for 1) secondary collimator (jaw) setting, 2) blocked field (multi-leaf collimator) setting, 3) Khan's theory using a 5 cm mini-phantom, and 4) Khan's theory using a 10 cm mini-phantom. Dose variations in each method of calculation were as follows: 1) +0.3 to +10.2% (mean, +2.0 to +3.2%) , 2) -2.3 to 0.0% (mean, -0.8 to -0.6%), 3) 0.0 to +1.5% (mean, +0.1 to +0.3%), 4) 0.0 to +1.4% (mean, -0.1 to +0.1%).

Original languageEnglish
Pages (from-to)1604-1610
Number of pages7
JournalNihon Hoshasen Gijutsu Gakkai zasshi
Volume60
Issue number11
Publication statusPublished - 2004
Externally publishedYes

Fingerprint

Hepatocyte Growth Factor
X-Rays
Ions
Polystyrenes
Jaw
Water

Cite this

Accuracy of absorbed dose calculation and measurement of scatter factors with different depth of mini-phantoms using a pinpoint ionization chamber. / Oita, Masataka; Watanabe, Yoshiharu; Fujita, Katsuhisa; Furuya, Teruo; Nankumo, Junya; Oda, Makoto.

In: Nihon Hoshasen Gijutsu Gakkai zasshi, Vol. 60, No. 11, 2004, p. 1604-1610.

Research output: Contribution to journalArticle

Oita, Masataka ; Watanabe, Yoshiharu ; Fujita, Katsuhisa ; Furuya, Teruo ; Nankumo, Junya ; Oda, Makoto. / Accuracy of absorbed dose calculation and measurement of scatter factors with different depth of mini-phantoms using a pinpoint ionization chamber. In: Nihon Hoshasen Gijutsu Gakkai zasshi. 2004 ; Vol. 60, No. 11. pp. 1604-1610.
@article{ef12704ca5a74b21a829fe4cdbc33878,
title = "Accuracy of absorbed dose calculation and measurement of scatter factors with different depth of mini-phantoms using a pinpoint ionization chamber",
abstract = "The output factor of high-energy X-ray machines varies with collimation. According to Khan's theory, collimator and phantom scatter factors contribute to total scatter factor. For precise X-ray irradiation, the two factors need to be taken into consideration. To obtain proper factors, we made two original polystyrene cylindrical mini-phantoms. These phantoms are both 4 cm in diameter and have a pinpoint ion chamber placed at a depth of 5 cm and 10 cm, respectively. Using a 6 MV X-ray machine, collimator scatter factors were calculated for various field arrangements (i.e., field sizes ranging from 4 cm x 4 cm to 40 cm x 40 cm at isocenter). To determine if calculated values were appropriate, we measured point doses of 20 X-ray irradiation patterns using a Farmer-type ion chamber with a water equivalent phantom at depths of 5 cm and 10 cm, respectively. Two hundred MUs were irradiated to the above-mentioned depths for each field. Based on the measured doses, variations were obtained for four calculation methods. Accounting for 1) secondary collimator (jaw) setting, 2) blocked field (multi-leaf collimator) setting, 3) Khan's theory using a 5 cm mini-phantom, and 4) Khan's theory using a 10 cm mini-phantom. Dose variations in each method of calculation were as follows: 1) +0.3 to +10.2{\%} (mean, +2.0 to +3.2{\%}) , 2) -2.3 to 0.0{\%} (mean, -0.8 to -0.6{\%}), 3) 0.0 to +1.5{\%} (mean, +0.1 to +0.3{\%}), 4) 0.0 to +1.4{\%} (mean, -0.1 to +0.1{\%}).",
author = "Masataka Oita and Yoshiharu Watanabe and Katsuhisa Fujita and Teruo Furuya and Junya Nankumo and Makoto Oda",
year = "2004",
language = "English",
volume = "60",
pages = "1604--1610",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "11",

}

TY - JOUR

T1 - Accuracy of absorbed dose calculation and measurement of scatter factors with different depth of mini-phantoms using a pinpoint ionization chamber

AU - Oita, Masataka

AU - Watanabe, Yoshiharu

AU - Fujita, Katsuhisa

AU - Furuya, Teruo

AU - Nankumo, Junya

AU - Oda, Makoto

PY - 2004

Y1 - 2004

N2 - The output factor of high-energy X-ray machines varies with collimation. According to Khan's theory, collimator and phantom scatter factors contribute to total scatter factor. For precise X-ray irradiation, the two factors need to be taken into consideration. To obtain proper factors, we made two original polystyrene cylindrical mini-phantoms. These phantoms are both 4 cm in diameter and have a pinpoint ion chamber placed at a depth of 5 cm and 10 cm, respectively. Using a 6 MV X-ray machine, collimator scatter factors were calculated for various field arrangements (i.e., field sizes ranging from 4 cm x 4 cm to 40 cm x 40 cm at isocenter). To determine if calculated values were appropriate, we measured point doses of 20 X-ray irradiation patterns using a Farmer-type ion chamber with a water equivalent phantom at depths of 5 cm and 10 cm, respectively. Two hundred MUs were irradiated to the above-mentioned depths for each field. Based on the measured doses, variations were obtained for four calculation methods. Accounting for 1) secondary collimator (jaw) setting, 2) blocked field (multi-leaf collimator) setting, 3) Khan's theory using a 5 cm mini-phantom, and 4) Khan's theory using a 10 cm mini-phantom. Dose variations in each method of calculation were as follows: 1) +0.3 to +10.2% (mean, +2.0 to +3.2%) , 2) -2.3 to 0.0% (mean, -0.8 to -0.6%), 3) 0.0 to +1.5% (mean, +0.1 to +0.3%), 4) 0.0 to +1.4% (mean, -0.1 to +0.1%).

AB - The output factor of high-energy X-ray machines varies with collimation. According to Khan's theory, collimator and phantom scatter factors contribute to total scatter factor. For precise X-ray irradiation, the two factors need to be taken into consideration. To obtain proper factors, we made two original polystyrene cylindrical mini-phantoms. These phantoms are both 4 cm in diameter and have a pinpoint ion chamber placed at a depth of 5 cm and 10 cm, respectively. Using a 6 MV X-ray machine, collimator scatter factors were calculated for various field arrangements (i.e., field sizes ranging from 4 cm x 4 cm to 40 cm x 40 cm at isocenter). To determine if calculated values were appropriate, we measured point doses of 20 X-ray irradiation patterns using a Farmer-type ion chamber with a water equivalent phantom at depths of 5 cm and 10 cm, respectively. Two hundred MUs were irradiated to the above-mentioned depths for each field. Based on the measured doses, variations were obtained for four calculation methods. Accounting for 1) secondary collimator (jaw) setting, 2) blocked field (multi-leaf collimator) setting, 3) Khan's theory using a 5 cm mini-phantom, and 4) Khan's theory using a 10 cm mini-phantom. Dose variations in each method of calculation were as follows: 1) +0.3 to +10.2% (mean, +2.0 to +3.2%) , 2) -2.3 to 0.0% (mean, -0.8 to -0.6%), 3) 0.0 to +1.5% (mean, +0.1 to +0.3%), 4) 0.0 to +1.4% (mean, -0.1 to +0.1%).

UR - http://www.scopus.com/inward/record.url?scp=16644362187&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=16644362187&partnerID=8YFLogxK

M3 - Article

VL - 60

SP - 1604

EP - 1610

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 11

ER -