A Study of uncertainty factors in cross calibration of dosimeter for diagnostic X-rays

Hirofumi Yagi, Hitoshi Kubo, Masataka Oita, Tokue Kuroda

Research output: Contribution to journalArticle

Abstract

Although patient exposure has been increasing in recent years, few institutions have dosimeters and are able to ascertain patients' exposure dose. Internationally, however, it is necessary to adopt safety levels for patient exposure doses, and guidance levels have been introduced. Therefore, the need for measurement in areas of x-ray diagnosis has been increasing. As a result, several societies concerned with radiation dose have been endeavoring to establish a calibration system of radiation measurement and a dosimeter calibration system, which are the basics of radiation protection. Ten regional centers for standardization of doses in x-ray diagnosis were established and have begun trials relating to dosimeter cross calibration. Our institution, as one of these centers, has instituted a trial. In this study, the cross-calibration field, the reliability of the cross-calibration skill of our regional center, and the standard uncertainty of cross calibration were investigated. As a consequence of the investigation, it was determined that our cross-calibration field follows the protocol of the Japanese Society of Radiological Technology, the difference between calibration factor/cross calibration factor obtained by JQA and our regional center is within 2.5%, and the expanded uncertainty of our cross calibration is about 7.2% (k=2).

Original languageEnglish
Pages (from-to)1006-1013
Number of pages8
JournalNihon Hoshasen Gijutsu Gakkai zasshi
Volume61
Issue number7
Publication statusPublished - 2005
Externally publishedYes

Fingerprint

Radiography
Calibration
Uncertainty
Radiologic Technology
X-Rays
Radiation Dosimeters
Radiation
Radiation Protection
Patient Safety

Cite this

A Study of uncertainty factors in cross calibration of dosimeter for diagnostic X-rays. / Yagi, Hirofumi; Kubo, Hitoshi; Oita, Masataka; Kuroda, Tokue.

In: Nihon Hoshasen Gijutsu Gakkai zasshi, Vol. 61, No. 7, 2005, p. 1006-1013.

Research output: Contribution to journalArticle

Yagi, Hirofumi ; Kubo, Hitoshi ; Oita, Masataka ; Kuroda, Tokue. / A Study of uncertainty factors in cross calibration of dosimeter for diagnostic X-rays. In: Nihon Hoshasen Gijutsu Gakkai zasshi. 2005 ; Vol. 61, No. 7. pp. 1006-1013.
@article{ae35ef68753a48fd954f9807447115c3,
title = "A Study of uncertainty factors in cross calibration of dosimeter for diagnostic X-rays",
abstract = "Although patient exposure has been increasing in recent years, few institutions have dosimeters and are able to ascertain patients' exposure dose. Internationally, however, it is necessary to adopt safety levels for patient exposure doses, and guidance levels have been introduced. Therefore, the need for measurement in areas of x-ray diagnosis has been increasing. As a result, several societies concerned with radiation dose have been endeavoring to establish a calibration system of radiation measurement and a dosimeter calibration system, which are the basics of radiation protection. Ten regional centers for standardization of doses in x-ray diagnosis were established and have begun trials relating to dosimeter cross calibration. Our institution, as one of these centers, has instituted a trial. In this study, the cross-calibration field, the reliability of the cross-calibration skill of our regional center, and the standard uncertainty of cross calibration were investigated. As a consequence of the investigation, it was determined that our cross-calibration field follows the protocol of the Japanese Society of Radiological Technology, the difference between calibration factor/cross calibration factor obtained by JQA and our regional center is within 2.5{\%}, and the expanded uncertainty of our cross calibration is about 7.2{\%} (k=2).",
author = "Hirofumi Yagi and Hitoshi Kubo and Masataka Oita and Tokue Kuroda",
year = "2005",
language = "English",
volume = "61",
pages = "1006--1013",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "7",

}

TY - JOUR

T1 - A Study of uncertainty factors in cross calibration of dosimeter for diagnostic X-rays

AU - Yagi, Hirofumi

AU - Kubo, Hitoshi

AU - Oita, Masataka

AU - Kuroda, Tokue

PY - 2005

Y1 - 2005

N2 - Although patient exposure has been increasing in recent years, few institutions have dosimeters and are able to ascertain patients' exposure dose. Internationally, however, it is necessary to adopt safety levels for patient exposure doses, and guidance levels have been introduced. Therefore, the need for measurement in areas of x-ray diagnosis has been increasing. As a result, several societies concerned with radiation dose have been endeavoring to establish a calibration system of radiation measurement and a dosimeter calibration system, which are the basics of radiation protection. Ten regional centers for standardization of doses in x-ray diagnosis were established and have begun trials relating to dosimeter cross calibration. Our institution, as one of these centers, has instituted a trial. In this study, the cross-calibration field, the reliability of the cross-calibration skill of our regional center, and the standard uncertainty of cross calibration were investigated. As a consequence of the investigation, it was determined that our cross-calibration field follows the protocol of the Japanese Society of Radiological Technology, the difference between calibration factor/cross calibration factor obtained by JQA and our regional center is within 2.5%, and the expanded uncertainty of our cross calibration is about 7.2% (k=2).

AB - Although patient exposure has been increasing in recent years, few institutions have dosimeters and are able to ascertain patients' exposure dose. Internationally, however, it is necessary to adopt safety levels for patient exposure doses, and guidance levels have been introduced. Therefore, the need for measurement in areas of x-ray diagnosis has been increasing. As a result, several societies concerned with radiation dose have been endeavoring to establish a calibration system of radiation measurement and a dosimeter calibration system, which are the basics of radiation protection. Ten regional centers for standardization of doses in x-ray diagnosis were established and have begun trials relating to dosimeter cross calibration. Our institution, as one of these centers, has instituted a trial. In this study, the cross-calibration field, the reliability of the cross-calibration skill of our regional center, and the standard uncertainty of cross calibration were investigated. As a consequence of the investigation, it was determined that our cross-calibration field follows the protocol of the Japanese Society of Radiological Technology, the difference between calibration factor/cross calibration factor obtained by JQA and our regional center is within 2.5%, and the expanded uncertainty of our cross calibration is about 7.2% (k=2).

UR - http://www.scopus.com/inward/record.url?scp=24944549166&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=24944549166&partnerID=8YFLogxK

M3 - Article

C2 - 16049414

AN - SCOPUS:24944549166

VL - 61

SP - 1006

EP - 1013

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 7

ER -