@inproceedings{44a92d0c8f34468bb9f4c3fcb8dfd2f4,
title = "A Measurement System for Circular Dichroism in Soft X-ray Absorption Using Helicity Switching by Twin Helical Undulators",
abstract = "A measurement system for circular dichroism (CD) of soft X-ray absorption has been developed using helicity switching of circularly polarized radiation by twin helical undulators at BL25SU of SPring-8. Difference of horizontal spot positions between the negative- and positive-helicity light at the sample position is within ±3μm. Measurement time for magnetic circular dichroism (MCD) of core absorption is shortened to less than 1/3 in comparison with an existing system of magnetic field switching, keeping the same data quality.",
author = "T. Muro and Y. Saitoh and H. Kimura and T. Matsushita and T. Nakatani and M. Takeuchi and T. Hirono and T. Kudo and T. Nakamura and T. Wakita and K. Kobayashi and T. Hara and K. Shirasawa and H. Kitamura",
note = "Publisher Copyright: {\textcopyright} 2004 American Institute of Physics.; 8th International Conference on Synchrotron Radiation Instrumentation ; Conference date: 25-08-2003 Through 29-08-2003",
year = "2004",
month = may,
day = "12",
doi = "10.1063/1.1757978",
language = "English",
series = "AIP Conference Proceedings",
publisher = "American Institute of Physics Inc.",
pages = "1051--1054",
booktitle = "Synchrotron Radiation Instrumentation",
}